15 Technical data and dimensions
15.1 General
mbar
bar
Pa
hPa
kPa
Torr | mm Hg
mbar
1
1 · 10
-3
100
1
0.1
0.75
bar
1000
1
1 · 10
5
1000
100
750
Pa
0.01
1 · 10
-5
1
0.01
1 · 10
-3
7.5 · 10
-3
hPa
1
1 · 10
-3
100
1
0.1
0.75
kPa
10
0.01
1000
10
1
7.5
Torr | mm Hg
1.33
1.33 · 10
-3
133.32
1.33
0.133
1
1 Pa = 1 N/m
2
Tbl. 10:
Conversion table: Pressure units
mbar l/s
Pa m
3
/s
sccm
Torr l/s
atm cm
3
/s
mbar l/s
1
0.1
59.2
0.75
0.987
Pa m
3
/s
10
1
592
7.5
9.87
sccm
1.69 · 10
-2
1.69 · 10
-3
1
1.27 · 10
-2
1.67 · 10
-2
Torr l/s
1.33
0.133
78.9
1
1.32
atm cm
3
/s
1.01
0.101
59.8
0.76
1
Tbl. 11:
Conversion table: Units for gas throughput
15.2 Technical data
Parameter
Value
Measuring range (air, N
2
)
5 × 10
-9
– 1000 hPa
Maximum pressure (absolute)
10000 hPa, limited to inert gases and temperatures < 55 °C
Measuring principle
Pirani/cold cathode
Accuracy
approx. ±30 % in the range of 1 × 10
-8
– 100 hPa
Repeatability
approx. ±5 % in the range of 1 × 10
-8
– 100 hPa
Gauge calibration (HV)
(Pirani measuring system)
at < 1 × 10
-4
hPa (while simultaneously pressing the button)
Gauge calibration (ATM)
(Pirani measuring system)
at atmospheric pressure
Tbl. 12:
Measured and pressure values
Parameter
Value
Output signal (measur-
ing signal)
Voltage range
approx. 0 V – approx. +10,5 V
Error signal
< 0.5 V (no supply)
> 9.5 V (Pirani measuring element de-
fective; broken filament)
Relation voltage-pressure
logarithmic, rise 0.6 V/decade
Output impedance
2 × 10 Ω
Load impedance (minimum load)
10 kΩ, short-circuit proof
Response time
pressure-dependent
p > 10
-6
hPa
approx. 10 ms
p = 10
-8
hPa
approx. 1 s
Technical data and dimensions