AFV-P Series High Performance Programmable AC Power Supply
76
B. Memory Loop
At the PROGRAMMABLE page, users are allowed to set the Memory Loop, and 50
Memory Sets are supported for simulating power line conditions and disturbance.
For example, when the start number of the Memory Loop is 2, the end number of
the Memory Loop is 6, and the Memory Loop times is 5, the Memory Loop will be
sequentially performed from the Memory Set 2 to the Memory Set 6 and repeated 5
times.
The procedures of setting the Memory Loop according to the example mentioned
above are given as below,
1. Press the icon
/
to enable the Memory Loop.
2. Press the icon
to set the value of 2, and press the icon
to confirm.
3. Press the icon
to set the value of 6, and press the icon
to confirm.
4. Press the icon
to set the value of 5, and press the icon
to confirm.
5. Press the output & reset button, then the Memory Loop is performed.
When the Memory Loop is performed, the following page will be shown on the touch
screen,
Figure 5.5 PROGRAMMABLE page when the Memory Loop is performed