AFV-P Series High Performance Programmable AC Power Supply
94
6
Calibration
The product provides a simple way to calibrate the product output and measurement
accuracy without opening cover. Users can perform the calibration according to the
procedures given as follows step by step. A voltage meter, a current meter and suita-
ble load are needed while performing the calibration procedures. Connections for
the instruments mentioned above please refer to the figure below.
Figure 6.1 Instrument connection for calibration
At the SYSTEM subpage 3 of the SETTINGS page, users can press the item
, and then use the virtual numeric keys to set the value of
8888, so as to enter into the CALIBRATION page. Please see the following figures,
Figure 6.2 Enter into the CALIBRATION page from the SYSTEM subpage 3