9.
Make sure the
New Base value
is written to the DSP.
Figure 6.11 Confirms the base value is updated
10.
Check whether the output current is correct with the voltmeter; if a deviation
still occurs, repeat step 3 to 9 again.
After completing the Voltage Calibration, press
Return
to go to the previous page for
additional calibration or press
General
to go to the General page for testing.
Notice
The New Base value can only be adjusted manually by authorized personnel.