6-1
6
Calibration
This product allows the user to calibrate the product output and the measurement
accuracy. Prepare a voltmeter, an ammeter and connecting wires beforehand for
calibration.
(a)
Voltmeter - open circuit
(b)
Ammeter
–
short circuit
Figure 6.1 Meter connection
To enter the Calibration page, press
Settings
at General page, press
Calibration
at
Setting page, and enter value 8888 via the virtual keyboard. Refer to the following
figures.
Notice
The meters and wires for calibration must match the output voltage rating and
output current rating of the product.