Rev 8/2014
PrecisionTemp, Inc • 11 Sunnybrook Drive • Cincinnati, Ohio, 45237 • 1-800-934-9690 •
fax 513-641-0733 46
SB-7 Automatic Calibration (AutoCal) Procedure
Model:
PT- 200FT
Software version
: 9097-A, 9097-B, 9097-C, 9097-D, 9097-E, 9097-F, 9097-G and
9099-A
Tools Required:
Needle-nose pliers
Process:
CAUTION:
During this procedure, the unit will be performing several self tests. Do not
attempt to use the dish machine during this period. The booster heater will increase the gas
flow from minimum to maximum over about 8 minutes. Be sure to monitor the unit for safe
operation.
1. Turn booster heater power switch off and remove front cover. Open electronic drawer.
Ensure that the software version on the microprocessor is one that is listed above.
2. From the front circuit board, remove the 2-pin jumper cap from
ACAL
(back jumper)
using pliers or fingers (see figure 200-14).
3. Lock the dish machine in rinse mode. (Rinse MUST stay on for the duration of this
process).
4. Place a magnet on the modulating gas line just right of the electrical coil and move it
slowly towards the gas solenoid.
5. Turn booster heater power switch on. Green LED should flash and burner ignite to high
burn. If red light flashes there is a problem that must be corrected before continuing.
See service code guide.
6. Remove the magnet SLOWLY from the gas line. Burner should return to low burn.
7. Over the next 8 minutes the unit will increase the gas flow to calibrate the valve. When
the test is complete, the burner will shut down and the red LED will flash rapidly.
8. Turn power switch off and wait 20 seconds.
9. Replace the 2-pin jumper cap on the circuit board and turn power switch back on.
10. Check the unit for normal operation. If operation appears normal the procedure is
complete. If the unit does not operate normally or gas flow varies excessively repeat the
procedure.
Kit Parts List:
None
Reference Illustrations:
Figure 200-14 View of Control Board
Содержание PT-200FT
Страница 11: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 11...
Страница 13: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 13...
Страница 14: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 14...
Страница 15: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 15...
Страница 16: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 16...
Страница 17: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 17...
Страница 18: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 18...
Страница 19: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 19...
Страница 23: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 23...
Страница 24: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 24...
Страница 25: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 25...
Страница 26: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 26...
Страница 27: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 27...
Страница 28: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 28...
Страница 29: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 29...
Страница 30: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 30...
Страница 31: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 31...
Страница 32: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 32...
Страница 33: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 33...
Страница 34: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 34...
Страница 42: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 42...
Страница 58: ...Rev 8 2014 PrecisionTemp Inc 11 Sunnybrook Drive Cincinnati Ohio 45237 1 800 934 9690 fax 513 641 0733 58...