SPECIFICATIONS
Model
6MFA1001
4MFA1001
1MFA3001
Flow Range
0-200 cc
0-1 lpm
0-3 lpm
Gas
Oxygen
Oxygen
Oxygen
Increments
20 cc
(starts at 20 cc)
.1 lpm
(starts at .1 lpm)
.125
from .125 to 1 lpm
.25 lpm
from 1 to 3 lpm
Accuracy
± 10 cc
from 0-100 cc
± 14 cc
from 100-200 cc
± .05 lpm
± .15 lpm
Max Flush Flow
Range
500 cc*
6 lpm*
20 - 40 lpm
Transport / Storage
Requirements
-40˚F (-40˚C) to
140˚ (60˚C)
* Restricted Max Flush Flow
NOTE: Storage / Transport outside the specified range may cause damage
to the flowmeter.
The effect on accuracy of flow due to variations in ambient temperature is
standard ac7.3% @ 0°C and -3.0% @ + 40°C.
Flowmeters calibrated at 50 psi (3.4 bar), 70°F (21°C), standard atmospheric
pressure.
International models are calibrated per specifications marked on Flow Tube.
Specifications are subject to change without prior notice.