PD6700-0L1 Field-Mount Loop-Powered Process Meter 

Instruction Manual 

 

18

 

Level Input Live Signal Calibration 
(

LVL CAL

The meter can be calibrated using a current source 
instead of scaling. This process will override 
previously programmed scaling of the level display. 

The use of a calibrated signal source is strongly  
recommended. 

 

 

Internal Calibration (

ICAL

There is 

no need to recalibrate

 the meter when 

first received from the factory.  
The meter is 

factory calibrated

 prior to shipment. 

The calibration equipment is traceable to NIST 
standards. 

 

The internal calibration 

is the meter’s master 

calibration that makes scaling the meter without a 
signal source possible. Use of a calibrated signal 
source is necessary to perform an internal calibration 
of the meter. Check calibration of the meter at least 
every 12 months. Incorrect calibration will affect the 
ability of the meter to properly read, scale, and display 
the input. 

 
Press 

MENU

, navigate to 

ADVANCE

 and press 

ENTER

 

to access the 

Advanced

 

Features Menu

. Press the 

UP 

arrow button to scroll to the 

Internal Calibration

 

menu (

ICAL

) and press 

ENTER

The meter displays 

4.000

. Apply a 4.000 mA signal 

and press 

ENTER

. The display flashes for a moment 

while the meter is accepting the signal. After the 
signal is accepted, the meter displays 

20.000

. Apply a 

20.000 mA signal and press 

ENTER

. The display 

flashes for a moment while the meter is accepting the 
signal. 

 

 

 

Notes: 

1.  The signal source must have a full-scale 

accuracy of 0.002% or better between 4 and 
20 mA in order to maintain the specified 
accuracy of the meter.  

2.  Allow the meter to warm up for at least 

15 minutes before performing the calibration 
procedure. 

 

Calibration Error Message (

SPN 

ERR

An error message indicates that the calibration 
process was not successful. After the error message 
is displayed, the meter will revert to the 

INPUT 1

 

calibration menu. The error message might be caused 
by inadvertently leaving the signal at the previous 
level or not maintaining the minimum span. Press the 

MENU

 button to cancel the current calibration process 

if necessary. 

Level Input Multipoint 
Linearization (

MULTIPT

This menu enables multipoint linearization for scaling 
and calibrating of the level display. 

Setting 

MULTIPT

 to 

ENABLE

 allows the level display to 

be scaled or calibrated using up to 32 points. See 

Scaling the Meter 

(

SCaLe

) on page 

13

 and 

Level Input 

Live Signal Calibration 

(

LVL CAL

) on page 

18

, to 

include 

a Number of Points

 (

NO PTS

) parameter 

before entering 

Input 1

. The 32-point linearization is 

used to linearize the display for non-linear signals. 
 

 

PD6700L Displaying Height in Feet and Volume  

(Using Multi-Point Linearization Feature) in Gallons 

in a Round Horizontal Tank. 

Information (

INFO

The 

Information

 menu shows the software 

identification number and version number. To 
determine the software version of a meter: 
Go to the 

Information

 menu (

INFO

) and press 

ENTER

 

button. 
Continue pressing 

ENTER

 to scroll through the 

software release number and software version. 
Following the information display, the meter will exit 
the 

Advanced

 

Features

 menu and return to run mode. 

Содержание vantageview Safe Touch PD6700-0L1

Страница 1: ...r Display Display Mountable at 0 90 180 270 SafeTouchTM Through Window Button Programming Display Input in Two Different Scales Simultaneously Great for Level Applications 20 Segment Level Indicator B...

Страница 2: ...Precision Digital Corporation All rights reserved Introduction The PD6700 0L1 is a rugged field mount loop powered meter fully featured for demanding applications in harsh environmental conditions Th...

Страница 3: ...ing the Meter SCaLe 13 Minimum Input Span 13 Scale Error Message SPN ERR 13 Scaling the Tank Level Indicator PERCENT 13 Configuring the Display DISPLAY 13 Setting the Tag TAG 14 Setting Up the Passwor...

Страница 4: ...imensions Side Cross Section View 6 Figure 3 Connector Board 7 Figure 4 Connections without Backlight 8 Figure 5 Connections with Loop Powered Backlight 8 Figure 6 Connections with Externally Powered...

Страница 5: ...ntal Operating temperature range 40 to 75 C Storage temperature range 40 to 75 C Relative humidity 0 to 90 non condensing Connections Screw terminals accept 12 to 22 AWG wire Enclosure NEMA 4X IP65 pl...

Страница 6: ...e the meter from box Inspect the packaging and contents for damage Report damages if any to the carrier If any part is missing or the meter malfunctions please contact your supplier or the factory for...

Страница 7: ...ght from external supply OUTPUT NPN open collector output positive OUTPUT NPN open collector output negative RESET Contact closure alarm acknowledge pull up to 3 VDC RESET Contact closure alarm acknow...

Страница 8: ...e to two terminals labeled RESET Connect to a contact closure source such as a relay or a pushbutton as shown in Figure 7 Figure 7 External Alarm Reset Acknowledge Connections Open Collector Output Co...

Страница 9: ...rd in the base of the enclosure SafeTouch Button Operation To actuate a button press and remove one finger to the window directly over the marked button area Remove finger to at least 4 inches away fr...

Страница 10: ...the Right arrow button to reset the maximum or minimum value while it is being displayed see Up Display Button below Press the Right arrow button to move to the next digit or decimal position during p...

Страница 11: ...play Enter Bottom Display menu TAG Tag Display a custom unit or tag VOLUME Volume Display volume VOL TAG Volume Tag Display volume and custom tag PCT HT Percent Height Display percent height PCT TAG P...

Страница 12: ...o the previous menu selections Setting Numeric Values The numeric values are set using the RIGHT and UP arrow buttons Press RIGHT arrow to select next digit and UP arrow to increment digit The digit b...

Страница 13: ...nearization MULTIPT menu selection prior to scaling and calibration of the meter see Advanced Features Menu ADVANCE on page 15 Scaling the Tank Level Indicator PERCENT The display includes a 20 segmen...

Страница 14: ...recorded in the space provided Model Serial Number Password __ __ __ __ __ Making Changes to a Password Protected Meter If the meter is password protected the meter will display the message LOCKED wh...

Страница 15: ...Menu Access the Advanced Features Menu by pressing ENTER at the ADVANCE menu in the Main Menu defined on page 11 The Advanced Features Menu is used to select 1 Open collector output configuration OUTP...

Страница 16: ...NPUT 1 Input 1 Set volume input 1 on the level display DSPLY 1 Display 1 Set volume display 1 INPUT 2 Input 2 Set volume input 2 on the level display DSPLY 2 Display 2 Set volume display 2 Display Par...

Страница 17: ...etween its normal display and ALARM until the alarm condition is cleared The alarm status will show on the display even if the output is not wired Input Signal Filter FILTER The noise filter is availa...

Страница 18: ...ve a full scale accuracy of 0 002 or better between 4 and 20 mA in order to maintain the specified accuracy of the meter 2 Allow the meter to warm up for at least 15 minutes before performing the cali...

Страница 19: ...s displayed will reset that value to the current display value Press the DISPLAY button a third time and the meter will display LOOP MA on the lower display and the measured input current in milliamps...

Страница 20: ...e Diagnostics menu and consult the recommendations described in the following table Troubleshooting Tips Symptom Check Action No display or faint display Check input signal connections Perform hard re...

Страница 21: ...tion or increment digit Cycle through maximum minimum and mA display mode ENTER Accept selection value and move to next selection Acknowledge alarms To enter the Advanced Features Menu hold the Menu b...

Страница 22: ...8990 www predig com LIM6700L_D SFT087 Ver 1 300 up 07 20 Contact Precision Digital Technical Support Call 800 610 5239 or 508 655 7300 Fax 508 655 8990 Email support predig com Sales Support Call 800...

Отзывы: