Loop Leader PD6620 Series Loop-Powered Rate/Totalizer Instruction Manual
47
Troubleshooting
Due to the many features and functions of the meter, it’s possible that the setup of the meter does not
agree with what an operator expects to see.
If the meter is not working as expected, refer to the recommendations below.
Reset Meter to Factory Defaults
When the parameters have been changed in a way that is difficult to determine what’s happening, it might
be better to start the setup process from the factory defaults.
To load factory defaults:
1. Press the
Menu
button to enter
Programming Mode
.
2. Press the
Up-Arrow
button
twice
and press
Enter
to access the
Advanced
menu.
3. Press the
Up-Arrow
button and press
Enter
to access the
System
menu.
4. Press the
Right-Arrow
button and press
Enter
to access the
Default
menu.
5. Press
Enter
twice in quick succession. The meter will load default settings and restart.
Determining Software Version
To determine the software version of a meter:
1. Press the
Menu
button to enter
Programming Mode
.
2. Press the
Up-Arrow
button
twice
and press
Enter
to access the
Advanced
menu.
3. Press the
Up-Arrow
button and press
Enter
to access the
System
menu.
4. Press the
Up-Arrow
button
twice
and press
Enter
to access the
Info
menu.
5. Press the
Right-Arrow
button to cycle through the meter information. When done, press the
Menu
button to return to the previous menu.
SYSTEM
SYSTM
AOUTCAL
SYSTM
DEFAULT
LOAD?
DEFAULT
RESET
DFAULT?
Meter Loads
Default Settings
and Restarts