PD8-6060 Explosion-Proof Dual Analog Input Process Meter
Instruction Manual
14
Transmitter Power Supplies
Meter Powers Transmitters
One of the most useful standard features of the
PD8-6060 is its built-in isolated, 24 V @ 25 mA power
supply to power the transmitters. This feature saves
money by eliminating an external power supply and
also simplifies wiring by reducing the number of
devices in the loop. It can be configured for 5, 10, or
24 V (default) by means of a simple internal jumper.
This power supply is even available on meters that
are powered from DC power (24 V @ 25 mA). To use
an external power supply instead of the internal power
supply, simply make connections to different terminals
on the ProtEX-MAX.
The following diagram illustrates how to wire the
ProtEX-MAX so it will power the transmitters:
Meter Powers 4-20 mA Output
Not only can the ProtEX-MAX power the 4-20 mA
input signal, but an additional power supply of
24 V @ 25 mA is provided with the 4-20 mA output
option to power the 4-20 mA output.
Fuse Prevents Current Overload
Another very useful aspect of the ProtEX-MAX is that
the current input is protected against current overload
by a resettable fuse. The fuse limits the current to a
safe level when it detects a fault condition, and
automatically resets itself when the fault condition is
removed.