ProtEX-MAX PD8-6060 Explosion-Proof Dual-Input Meter Instruction Manual
31
Error Message (
Error
)
An error message indicates that the calibration or scaling process was not successful.
After the error message is displayed, the meter reverts to input 2 during calibration or scaling and to input
1 during internal calibration, allowing the appropriate input signal to be applied or programmed.
The error message might be caused by any of the following conditions:
1. Input signal is not connected to the proper terminals or it is connected backwards.
2. Wrong signal selection in
Setup
menu.
3. Minimum input span requirements not maintained.
4. Input 1 signal inadvertently applied to calibrate input 2.
Minimum Input Span
The minimum input span is the minimum difference
between input 1 and input 2 signals required to complete
the calibration or scaling of the meter.
Calibrating the Meter with External Source
To scale the meter without a signal source,
refer to
Scaling the Meter without a Signal
page 30
Warm up the meter for at least 15 minutes
before performing calibration to ensure
specified accuracy.
The meter can be calibrated to display the process variable in engineering units by applying the
appropriate input signal and following the calibration procedure.
The use of a calibrated signal source is strongly recommended to calibrate the meter.
Calibrating the Meter for Channel A (
CAL-A
)
Calibrating the Meter for Channel B (
CAL-B
)
Input Range
Input 1 & Input 2 Span
4-20 mA
0.15 mA
10 VDC
0.01 VDC