Load-change-Mode ansolute and relative:
Load-change-Mode absolute:
In this mode the load change must exceed a certain limit to be recognized as such.
Enter the minimum load change between measurements in g.
This setting is not possible for automatic value acquisition with time interval.
Load-change-Mode relative:
The percentage value defines both the range which must be left between measurements and
the limits within which the sample values must lie.
This mode allows the measurement of series with samples of the same weight, e.g. a series
with samples of 100 g, where the first sample is used as reference. When the next sample is
placed on the weighing pan, the weight must be below 50 g or above 150 g for a short time
in between (load /- 50%) and the weight of all subsequent samples must be in the
range between 50 g and 100 g (sample range +/- 50%).