390
System Settings
Precisa Gravimetrics AG
59
Temperature controlled:
The balance monitors the current ambient temperature. If the deviation from the ambient
temperature during the last adjustment is greater than the threshold, an adjustment is
triggered.
Delta temperature:
Threshold at which deviation triggers a new adjustment.
Time controlled:
Always triggers an adjustment at the set time.
Time:
Time when the time-controlled adjustment should take place.
AOS temp. Controlled:
The scale uses the AOS to monitor the current air temperature. If the deviation from the air
temperature during the last adjustment is greater than the threshold, an adjustment is
triggered.
AOS Delta temperature [° C]:
The threshold triggers a new adjustment if it deviates.
AOS humidity controlled:
The scale uses the AOS to monitor the current air humidity. If the deviation from the air
humidity during the last adjustment is greater than the threshold, an adjustment is
triggered.
AOS Delta humidity [%]:
The threshold triggers a new adjustment if it deviates.
AOS pressure controlled:
The scale uses the AOS to monitor the current air pressure. If the deviation from the air
pressure during the last adjustment is greater than the threshold, an adjustment is
triggered.
AOS Delta pressure [hPa]:
The threshold triggers a new adjustment if it deviates.
Содержание 350-8182-000
Страница 1: ...S W I S S M A D E 390 HF HA HE HM Operating Instructions 350 8182 000...
Страница 11: ...390 Your new balance Precisa Gravimetrics AG 10 1 6 Assembling Optional Click Click...
Страница 12: ...390 Your new balance Precisa Gravimetrics AG 11 45 Click...
Страница 13: ...390 Your new balance Precisa Gravimetrics AG 12...
Страница 26: ...390 Discover the Capabilities of the series 390 Precisa Gravimetrics AG 25...