Model 3710 Plasma Control Console
Operator’s
Manual
_____________________________________________________________________________________________________________________
Issued: 03/11
Page
30
cable (P/N
5007777
) to the console’s JY connector and then to the JX
connector on the E-Stop-equipped system component.
3. When there are no more units with open JX connections, place the E-
Stop circuit completion plug in the JY connector of the last unit (plug
is supplied with the E-stop box). If the console is the only system
component connected to the E-stop box, the E-Stop circuit completion
plug will be placed in the JY connector on the console.
4. Connect a user-supplied cable from a switch closure to console’s
DUST COLLECTOR INTERLOCK connector.
Dust collector interlock
is designed to prevent system operation without
the booth exhaust system in operation. The dust collection system must be
on before the console can be energized. An air-flow switch in duct work,
connected to the console DUST COLLECTOR INTERLOCK connector is
one of the means to satisfy this safety requirement. For more information
on this feature, contact our service department.
Console will not power up if interlock is not connected to air flow
sensor or jumpered out.
B.
Compressed Air Connections
1.
Attach hose # 4 (P/N
5005064
) to console AIR IN fitting, then to a
clean, dry, oil-free compressed air supply.
C.
Supply Gas Connections
1.
Attach hose # 5A (first of three P/N
5005071
) to console
PRIMARY IN fitting on manifold, then to primary gas supply.
2.
Attach hose # 5B (second of three P/N
5005071
) to console
SECONDARY IN (inert gas) fitting on manifold, then to inert
secondary gas supply.
3.
Attach hose # 5C (third of three P/N
5005071
) to console
CARRIER GAS IN (inert gas) fitting on manifold, then to inert
carrier gas supply.
JY