![background image](http://html1.mh-extra.com/html/pr-electronics/9113ax/9113ax_product-manual_1593052050.webp)
10
Version No. V6R0
Safety Manual
9113 TEMPERATURE / mA CONVERTER
14 Safe parameterisation - user responsibility
14.1 Safety-related
configuration
parameters
14.1.1 Parameters related only to Channel 1
Name
Function
CH1.TYP
Selected input type:
TEMP = Temperature
CURR = Current
I.RANGE
Selected fixed input range for current measurements (for CH1.
TYP = CURR):
0_20 = 0...20 mA (no sensor error detection!)
4_20 = 4...20 mA
CONNEC
Selected sensor connection type for RTD measurements (for
SENSOR = Ni or Pt):
2W = 2-wire
3W = 3-wire
4W = 4-wire
If 2W or 3W is selected, the end user must ensure that the
applied sensor wiring does not introduce failures exceeding the
requirements for the safety application.
UNIT
Selected temperature unit (for CH1.TYP = TEMP):
°C = degrees Celsius
°F = degrees Fahrenheit
SENSOR
Selected temperature sensor type (for CH1.TYP = TEMP):
TC = Thermocouple
Ni = Ni RTD sensor
Pt = Pt RTD sensor
Pt.TYPE
Pt sensor type (for SENSOR = Pt):
10 = Pt10
20 = Pt20 (No short circuit detection!)
50 = Pt50
100 = Pt100
200 = Pt200
250 = Pt250
300 = Pt300
400 = Pt400
500 = Pt500
1000 = Pt1000
}