S3x-HB-1-11
45
5
External control interface command reference
The Sentinel3 system supports full remote command emulation over either USB or LAN interfaces.
The remote command set allows configuration and control of the Sentinel 3 system. These commands
are enacted by the Sentinel 3 Controller, which configures transmitter heads via the receiver modules.
Figure 54: Ethernet to GPIB external adapter
The commands and responses are text-based and not bound to any particular higher-level protocol.
Ethernet and USB-serial physical connections are provided with provision for GPIB connection using an
external adapter.
Figure 55: USB to GPIB external adapter
The existing Sentinel IIsc
command set is based on the product’s physical topology in which each
individual receiver module is paired with a single transmitter head with up to four RF inputs command
structure. As the Sentinel 3 product has a significantly different architecture it inherently disallows
backward-compatibility with the old command set. Where feasible, equivalent commands have been used
to ease the migration path for users; this equivalence is detailed later in this section.
5.1 Ethernet connection
Sentinel 3 is configured as a DHCP client. This requires that it is connected to a network supporting a
DHCP server. Typically this is provided by a LAN router. Alternatively, if no LAN is available, a local
router with a DHCP server can be used to connect Sentinel 3 to a remote controller PC. No uplink or
internet connection is required. Most low-cost SOHO routers support this functionality.
To create a peer-to-peer Ethernet connection, a cross-over Ethernet cable will be required if the PC host
does not support MDIX, otherwise any Ethernet cable may be used. The IP address on the host PC and
Sentinel 3 must be manually configured to share the same subnet, conventionally 192.168.0 for private
networks, but with different final IP numbers in the range 1-255. For example, IP addresses 192.168.0.10
and 192.168.0.20 would be suitable for the two devices. Where required, the subnet mask should be set
to 255.255.255.0. By default, Sentinel 3 uses port address 65000 to avoid clashing with other Ethernet
services.