8
HumiTherm Plus for PC
User Manual
Parameter Description
Settings
(Default Value)
0 to 23 for Hour
0 to 59 for Min
(Default : NA)
RTC TIME >>
Set current clock time in Hrs:Min (24 Hours format).
1 to 31
(Default : NA)
RTC DATE >>
Set current calendar date.
1 to 12
(Default : NA)
RTC MONTH >>
Set current calendar month.
2000 to 2099
(Default : NA)
RTC YEAR >>
Set current calendar year.
1 to 127
(Default : 1)
DEVICE ID >>
(Applicable for Serial Communication)
Unique numeric code assigned to the indicator for identification by
the remote PC. Set the value as required by the remote PC.
1000 to 1999
(Default : 123)
CHANGE PASSWORD >>
The Controller is shipped from the factory with a default password
(123) for accessing the parameters reserved for Supervisor.
However, if required the password can be changed by setting the
new value for this parameter.
(The new password replaces the old password. That is, the old
password is no longer valid. it is user’s responsibility to memorize
the password.)
1 to 128
(Default : 1)
GSM MACHINE ID >>
(Available only if the controller is supplied with GSM Module
Version)
This parameter can be used to assigned a unique ID to the
machine (chamber) to identify the source of SMS alert.
%RH ZERO OFFSET >>
This value is algebraically added to the measured RH Value to
derive the final Value that is displayed and compared for alarm /
control. Use this value to nullify any known constant error.
Final Value = Measured Value + Offset
This value is algebraically added to the measured Temperature
Value to derive the final Value that is displayed and compared for
alarm / control. Use this value to nullify any known constant error.
Final Value = Measured Value + Offset
TEMP. ZERO OFFSET >>
-50.0 to 50.0
(Default :
)
0.0
-50.0 to 50.0
(Default :
)
0.0