TARGAS-1 Operation Manual V. 1.02
92
Custom – Start Process Menu
Back Button
Returns to the Processes Menu.
Right Arrow Button
Confirms and starts the Custom process by displaying the Custom – Volume
and Area Settings menu (Step 2).
Custom – Volume and Area Settings (Step 2)
The volume and area settings are defined in this screen.
Important Note
. You must remember to account for any volume changes in order for the instrument to
properly calculate assimilation rates. If using collars, please make sure that you adjust the volume
accordingly.
Custom Step 2 Menu
Volume Button
Sets the volume of the system. Entries between 10 and 100,000 ml are
permitted.
Area Button
Sets the surface area of the chamber in cm
2
. Entries between 1 and 10,000
cm
2
are permitted.
Volume/Area Ratio
This value is automatically updated when there is a change to either the
Volume or Area values.
Stop Button
Stops the Custom process and returns to the Processes Menu.
Right Arrow Button
Continues to the Custom – Termination Settings menu (Step 3).
Custom – Termination Settings (Step 3)
The Termination settings (DT, DC) and Delay are defined in this screen.