TARGAS-1 Operation Manual V. 1.02
132
Where
‘XC’ are the Co2 Irga zeros
‘XH’ are the H2o Irga zeros
Z
Character
“Z” Command (Do a Zero)
This command initiates a zero. During a zero operation, the TARGAS-1 transmits
a Z string every second until the end of the zero operation. No “M” strings are sent
during this time. This command can be initiated by the user or by the TARGAS-1
itself.
Example:
Sent from Host: Z<CR>
Received from TARGAS-1:
+<CR>
Z, 025<CR>
Z, 024<CR>
…
A zero has been initiated by the user. In this example, 025 and 024 indicate the
number of seconds remaining in the zero operation.
Get/Set Parameter Values Table
Get/Set Parameter Values
Number
(X)
Parameter
Name
Description
Valid Values
Default Value
1
Zero Type
See Zero Settings
section. The values are:
1=Automatic, 2=User Set,
3=Manual
1 - 3
1 (Automatic)
2
Zero Time (min)
See Zero Settings
section.
1 - 40
20
3
Average Limit (ppm)
See Averaging Settings
section.
0 - 100
30
4
Alarm Low (ppm)
See Alarm Settings
section.
0 - 100000
0
5
Alarm High (ppm)
See Alarm Settings
section.
0 - 100000
2000
6
Alarm Sound
See Alarm Sound
section.
0=Off; 1=On
0 (Off)
7
CO
2
Concentration (ppm)
Not used in TARGAS
n/a
n/a
8
TARGAS ID
See TARGAS ID section.
0-9
1
9
Pump Power (%)
Not used in TARGAS
n/a
n/a