![background image](http://html1.mh-extra.com/html/pp-systems/egm-5/egm-5_operation-manuals_1592669081.webp)
EGM-5 Operation Manual V. 1.04
81
Calibration
O
2
Conc Button
O
2
concentration (%) of the calibration reference gas.
Entering a value will begin the Calibration Process
Actual O
2
The current O
2
reading based on the Scale Factor
O
2
Scale Factor
The current Scale Factor being used
Stop Button
Returns to the Calibration Menu.
Connecting Calibration Gas to the EGM-5
During calibration, the signal from the O
2
sensor is referenced against a gas of known O
2
concentration.
For greater accuracy, reference gas mixtures should be accurate to at least +/- 1% and traceable to NIST
standards.
Follow these instructions to calibrate the EGM-5 O2 sensor:
1. To avoid excess pressure in the sample cell and damage to the analyzer, you must place a T-fitting
between the reference gas source and the
GAS IN
port on the EGM-5. One end of the T-fitting should
be left open in order to relieve excess pressure, as follows:
Air with
O
2
GAS
Cylinder
Pressure
Regulator
Shut-off
Valve
“T”
To vent
Gas
~ 250ml/min
Gas In
Port