Page 3 of 8
11.
When all connections are complete and leak tested with an
oxygen compatible leak test solution, fasten the sleep lab front
panel to the back box with the four #8-32 x ¾” screws provided.
12.
On the front panel, label each flowmeter in the space provided
with the sleep room it controls (label not provided). See
Appendix A.
Mounting Screws
Miss-connection of the gases could lead to serious
or fatal injury to patients. Following installation,
DISS connections must be tested for cross-connection (per appropriate
standards) to ensure that the intended services are correctly connected to
the appropriate service lines.
Make certain the labeling coincides with the gas
service, and areas controlled by the Sleep Lab
Panel and that it is easily read.
Maintenance
1.
Clean the exterior of the stainless steel Sleep Lab Panel with a
glass cleaner. Strong solvents will damage the etched screened
printing on the sleep lab panel.
2.
Never use glass cleaner on the flowmeters.
3.
If it is necessary to clean the flowmeters wipe them clean with
a cloth dampened with water
4. Wipe dry with a clean dry cloth.
Flowmeters
Accuracy: ± 10% of the indicated flow.
Operating Temperature: 69°F (20.6°C) to 71F° (21.6° C)
Storage Temperature: -40°F (40°C) to 140°F (60° C)
Max Humidity: 95 %
Note:
Storage / Transport outside the specified range may cause
damage to the flow meter.
Specifications are subject to change without notice.
Operating Instructions
Note:
Tri-Tech Medical, Inc. recommends the use of kink proof
patient tubing.
1.
Be sure Dial Flowmeter is in the “OFF” position.
2.
Connect the Dial Flowmeter to a 50 psi (3.4 bar) Oxygen Source.
3.
Set the flow by aligning the Indicating Pointer
▼
to the desired
flow.
Note:
The accuracy of flow will be adversely affected by any variation
of inlet pressure, ambient temperature or outlet resistance from
rated values.
Read this user manual before installing or
operating the Dial Flowmeter.
There is NO FLOW between settings. The Indicating Pointer
▼
must point
to a specific number on the dial to obtain flow.
ONLY use appropriate indexed fittings to connect
the Dial Flowmeter to appropriate source. Use
Oxygen connections for oxygen flowmeters and use Air connections for Air
flowmeters.
Cleaning/Maintenance
1.
Disconnect all connections before cleaning.
2.
Clean exterior surfaces of the Dial Flowmeter with a cloth
dampened with water.
3. Wipe dry with a clean cloth.
4.
Store the Dial Flowmeter in a clean area free from grease, oil
and other sources of contamination.
DO NOT immerse the Dial Flowmeter in any kind
of liquid. This will cause damage to the Dial
Flowmeter and will void the warranty.
This Dial Flowmeter is not field serviceable.