However, the performance of PowerDot
PD-01M device can still affected by radio
frequency fields originating from other
sources.
For more information about EMC
emissions and immunity, contact
Smartmissimo Technologies.
The device should not
be used adjacent to
or stacked with other
equipment. If adjacent or
stacked use is necessary,
the device should be
observed to verify
normal operation in the
configuration in which it
will be used.
ELECTROMAGNETIC
COMPATIBILITY
(EMC)
PowerDot PD-01M is designed to be
used in home healthcare environments
in accordance with the EMC safety
standard IEC 60601-1-2 (4th Edition) and
with limitations, defined by the warnings
and precautions in this manual (e.g.
operation near RFID emitters, working
microwave ovens, etc.).
Examples of home healthcare
environment include restaurants,
cafes, shops, stores, markets, schools,
churches, libraries, outdoors (streets,
sidewalks, parks), domiciles (residences,
homes, nursing homes), vehicles (cars,
buses, trains, boats, planes, helicopters),
train stations, bus stations, airports,
hotels, hostels, pensions, museums,
theaters.
PowerDot PD-01M is designed to support
anticipated disturbance originating from
electrostatic discharge, magnetic fields
for the power supply or radio-frequency
emitters.
PowerDot battery charging
performance might
degrade in environments
with frequent voltage
interruptions (when
charging from a wall
adapter). To support
consistent and reliable
charging, the usage of UPS
is highly recommended,
if operating in such
environments.
The following device function is
considered essential to the safety of
the user: ability to maintain consistent
stimulation intensity (amplitude), pulse
frequency and pulse waveform (both
shape and width).
In case if the essential performance is
lost or degraded due to electromagnetic
disturbances, stimulation safety and
effectiveness can be compromised.
Whenever the user realizes unexpected
change in any of stimulation parameters,
it is advised to terminate the stimulation
session immediately by using one of the
methods provided in the Terminating
Stimulation section.
The use of accessories,
transducers and cables
others than those
specified or provided by
the manufacturer of this
equipment could result in
increased electromagnetic
emissions or decreased
electromagnetic immunity
of this equipment and
result in improper
operation.
Portable RF communica-
tions equipment (including
peripherals such as an-
tenna cables and external
antennas) should be used
no closer than 30 cm (12
inches) to any part of the
PowerDot PD-01M, includ-
ing cables specified by the
manufacturer. Otherwise,
degradation of the perfor-
mance of this equipment
could occur.
Содержание Duo
Страница 1: ...1 INSTRUCTION MANUAL PowerDot PD 01M Rev 1 0b 20 Dec 2017 ...
Страница 15: ...28 29 ABS OBLIQUES PECS ...
Страница 16: ...30 31 DELTOIDS BICEP ...
Страница 17: ...32 33 TRICEP FOREARM ...
Страница 18: ...34 35 LATS TRAPS ...
Страница 19: ...36 37 LOWER BACK ADDUCTORS ...
Страница 20: ...38 39 QUADS HAMSTRINGS ...
Страница 21: ...40 41 CALVES GLUTS ...
Страница 22: ...42 43 ELBOW SHOULDER ...
Страница 23: ...44 45 MIDDLE BACK BELOW NECK ...
Страница 24: ...46 47 KNEE TIBIA ...
Страница 25: ...48 49 ANCLE FOOT ...
Страница 26: ...50 51 IT BAND ...
Страница 35: ...68 ...