PostProcess Technologies, Inc. • www.postprocess.com • REV C | page 6
1. OPERATION
1.2 Auto Operations
If there is not much variance in processing settings, the CENTI can be set up to run in AUTO Mode after turning it
on via the
ON/STANDBY
button. Pressing the
AUTO
button starts the CENTI automatic operating mode. When
this mode is activated, the CENTI will perform the following actions: (Keep in mind that the unit will use the last
set of parameters that were entered via the
Fn
button.)
•
The heater will turn on to bring the detergent up to the Set Temperature. See
Section 1.3.2 – Setting the Target Temperature
.
•
Once the set temperature is reached the Ultrasonics are activated. See
Section 1.3.4 – Setting the Ultrasonic Power
.
•
Once the ultrasonics timer reaches “0” minutes the Auto Cycle is complete. See
Section 1.3.3 – Setting the Ultrasonic Timer
.
1.3 Using the Control Screen
The
Control
Screen shown below is the operating screen to access all CENTI functions. The screen displays all of
the current operating metrics during a cycle.
Figure 3.
Control Screen
NOTE
If the detergent in the tank does not reach the Set Temperature within the first 120 minutes, the Auto Cycle will
shut down. If this occurs, the screen will show the Auto icon flashing. If during an Auto Cycle, the
DEGAS
button
is pressed, the cycle will pause and the ultrasonics will restart after the Degas cycle.
See Section 1.4 - Degas
Operations
.