background image

7.

Press the 

EXECUTION 

button and the message 

Calib Drop # 1

appears on the readout (Photo I).

8.

Press the 

EXECUTION 

button once again and the message 

Raise

Weight

appears on the readout (Photo I). 

9.

As the weight is raised to the maximum height allowed (against the 
coupling), the message 

Drop Weight

will appear. 

Photo I 

10.

Drop the weight (after the weight hits the spring in the housing, catch the weight with one hand as it 
returns upward so as to halt the bouncing action). 

11.

After catching the weight, return it to it’s resting position within the housing. 

IMPORTANT – The weight must be allowed to come to rest on the spring in the housing between trials, to 
quiet the vibration of the spring.  Neglecting to do so may produce erroneous measurements. 

12.

After steps 10 and 11 have been performed, a message on the readout will give the percentage of the energy 
absorption. 

13.

Continue pressing the 

EXECUTION 

button to repeat for 

Calib Drop # 2

and

Calib Drop # 3

.   After 

Drop # 3, the average of the three drops will be shown on the readout. 

14.

Press the 

EXECUTION 

button once again, and either 

Calib Passes

or 

Calib Fails

will appear on the 

readout.

15.

If 

Calib Fails

appears, repeat the calibration steps on another (concrete) floor area.

Basket Height Measurement

4

1.

Place a piece of duct tape over the 
center of the rim at the furthest point 
from the backboard and hook the tester 
onto the rim over the tape (Photo J). 

2.

As   you   pull   downward,   the   hook 
should click into place as the spring- 
loaded detents engage the rim. 

3.

Allow the unit to come to rest in a 
plumb (vertical) position. 


Photo J 

Photo L

4.

Unscrew the  bottom  shaft  with  the  knurled  handle  (Photo  K). 
Lower this measuring shaft until it touches the floor (Photo L).

5.

Look for the scored line on the black anodized shaft – this line is 
used to check if the goal is at the required 10'-0" height. 

Photo K 

Содержание FAIR-COURT

Страница 1: ...N MAINTENANCE MANUAL FAIR COURT OFFICIAL BASKETBALL EQUIPMENT TESTING SYSTEM No ERTD2003NCAA SAVE THESE INSTRUCTIONS FOR FUTURE USE INST 00378 063 2011 PORTER ATHLETIC EQUIPMENT COMPANY ALL RIGHTS RES...

Страница 2: ...y and a new battery should be installed immediately Battery life is 20 40 hours depending on battery quality and use When in execute mode Drop Weight uses the most electricity and will drain the batte...

Страница 3: ......

Страница 4: ...give the percentage of the energy absorption 13 Continue pressing the EXECUTION button to repeat for Calib Drop 2 and Calib Drop 3 After Drop 3 the average of the three drops will be shown on the rea...

Страница 5: ...ure to avoid possible injuries or damage to the unit 3 Press the MODE button and the message Test Bskt 1 will appear on the readout Photo M 4 Press the EXECUTION button and the message Bskt 1 Drop 1 w...

Страница 6: ...r goal The final message will indicate whether the court has met the criteria for Rebound Range and Rebound Differences A Court PASS or Court FAIL message will appear 5 This concludes the test 6 The l...

Страница 7: ...PROBLEM Reading seem erroneous or unordinary SOLUTION If you are in a mode such as Calibration or Basket 1 or 2 you may cycle through and replace previous data SOLUTION If readings are still out of t...

Страница 8: ...rage 2 Backstop No 1 Passes Fails 3 Backstop No 2 Passes Fails 4 Difference Passes Fails NCAA CALIBRATION REQUIREMENT The rebound elasticity of any basket ring support system shall be within a 35 perc...

Отзывы: