Polimaster PM1701M Скачать руководство пользователя страница 12

 

12 

ambient background level or in slightly longer periods of time with an increasing 
ambient background level. However, the instrument will automatically recalibrate 
itself only if there are no alarms or abrupt background changes over the algorithm-
defined periods of time.  

Bear in mind that false alarms (audible or  light) are sporadic and are easily 

distinguished from true alarm signals which tend to be continuous or increase as the 
instrument moves closer to a radiation source.  

When a radiation source is detected, the operator may locate it. To locate a 

radiation source, hold the instrument at a distance 10 cm or closer from the scanned 
object, the scan velocity should be 10 cm per second or slower. As the instrument 
moves closer to the radiation source, the rate of alarm signals increase. When the 
alarm  signals reach the maximum rate pinpointing the source more closely is 
impossible without updating the background. Keeping the same distance to the 
scanned object press and hold the MODE button 12  or button 10 (Fig. 1) (if the 
headphones is used), until “CAL” is displayed to start the calibration sequence. The 
alarm threshold is automatically adjusted to the new baseline locating the source may 
be resumed. It may be necessary to repeat this procedure several times until a 
radiation source is located. 

 
2.2.4 Settings 

 

To enter this mode press and hold the MODE button 12 (Fig. 1) for more than 

4 seconds. Shortly press the MODE button to select a parameter to be set:  

 

check the preset or set a new value of coefficient n (number of meansquare 

deviations). The setting rage of the coefficient n is from 1 to 9.9 in 0.1 steps; 

 

check the status of the built-in  audible  alarm  and/or  headphones  audible 

alarm and change it (switch on/off), if this is enabled in the PC communication mode. 

 
2.2.4.1 Setting the Coefficient n  

 
To set the coefficient n press and hold the MODE button 12 (Fig. 1) for more 

than 4 seconds to enter the Settings mode. The LCD will display “CAL” message and 
then the preset value of coefficient n.  

 

 

 

Содержание PM1701M

Страница 1: ...PM1701M ALARMING RATEMETER Operation Manual...

Страница 2: ...Modes 10 2 2 1 Self Check 10 2 2 2 Background Calibration 10 2 2 3 Search Mode Detection and Location of Radiation Sources 11 2 2 4 Settings 12 2 2 4 1 Setting the Coefficient n 12 2 2 4 2 Switching O...

Страница 3: ...nstrument alerts the user by visual and audible signals about the presence of radiation source nearby All operation history is stored in the instrument s non volatile memory and may be downloaded to a...

Страница 4: ...e intrinsic error of count rate measurement temperature and humidity variations from normal to high temperature variations from normal to low 15 o C extreme values of supply voltage 40 10 10 PC commun...

Страница 5: ...the instrument off and follow instructions above to replace the battery 1 4 Design Control Buttons and Display The instrument Fig 1 is composed of measuring unit 1 attached to telescopic pole 4 The ma...

Страница 6: ...6 Figure 1 PM1701M The control buttons view A are used to 12 MODE Button turn the instrument on update background press and hold the button until the LCD displays CAL message...

Страница 7: ...itch off Vibr on headphones alarm in the setting mode LCD indication 15 LCD upper line count rate in count per second s 1 messages test CAL Aud etc coefficient n value in the setting mode Analog Scale...

Страница 8: ...ight alarm The 16 bit RISC based processor is used to test of the instrument upon power on process data control all operating modes display processed information on LCD control audible alarms monitor...

Страница 9: ...signals increases as the device moves closer to a radiation source The audible alarms can be switched on off in the PC communication mode or using the control buttons on the front panel of the instrum...

Страница 10: ...ally occurring background radiation The processor counts pulses coming from the detector for the preset time and the analog scale segments increase as the instrument calibrates The scale completely fi...

Страница 11: ...in the processor memory are updated every 0 25 s Every 0 25 second the current average counts are compared with the alarm threshold calculated during background calibration If the current average cou...

Страница 12: ...dating the background Keeping the same distance to the scanned object press and hold the MODE button 12 or button 10 Fig 1 if the headphones is used until CAL is displayed to start the calibration seq...

Страница 13: ...m with the control buttons provided that this mode is enabled in the parameters set in the PC communication mode If this mode is enabled the built in audible alarm is switched on off in the following...

Страница 14: ...omputer Install the Software PM17 _14XX supplied on CD Run SETUP EXE program using the installation instructions install doc autorun Help file and attached text document describe how to use the progra...

Страница 15: ...on off audible alarms enable disable switching on off the alarm using control buttons check or change coefficient n check or change the time and date check or change the logging intervals in which co...

Страница 16: ...ery replacement and regular check up of the instrument Routinely examine the instrument body remove any dust and decontaminate the instrument in situation of direct contact with radioactivity Wipe the...

Страница 17: ...st of the problems can be traced to low battery or poor quality battery Use only fresh high quality batteries There are no user serviceable parts in the instrument If repairs are required contact the...

Страница 18: ...f the assembly not more than 5 at 0 95 confidence probability 6 4 Barometer Least division 1 kPa Measurement range from 60 to 120 kPa 6 4 Thermometer Least division 0 1 C Measurement range from 10 to...

Страница 19: ...wait at least 60 s Then take 5 background readings Nbi s 1 and calculate b N s 1 by formula 5 1 5 1 i bi b N N 6 1 2 place the instrument on the dosimetric calibration assembly with 137 Cs gamma sour...

Страница 20: ...20 Compare main relative error to limits of permissible main error acc 35 If 35 the instrument is rejected as defective if 35 the instrument is considered ready for use...

Страница 21: ...r instruments should be packaged in hermetic compartments The instruments should be stored in the original manufacturer s package with the air temperature between 15 to 50 C and humidity less than 95...

Страница 22: ...period is 6 months since the date of manufacturer Quality Control Department accepting the instrument 8 4 Warranty and after warranty repairs are performed by the manufacturer or by a company authori...

Отзывы: