If the possible solutions suggested below do not address your problem, please call our
Customer Service department at 1-800-763-8423 and we will do everything possible to help.

NOTE: Never run the pump without water flowing through the lines! This will cause pump failure.

IF THE PUMP IS NOT RUNNING

Check the wall outlet to assure there is power to the unit. If the battery pack option is
being used, make sure the battery is fully charged. Confirm that all power cords are
properly plugged in and seated. 

IF THE THERAPY BLADDER IS NOT GETTING COLD

First: Is water flowing out of the top water return flow tube in the cooler?

If water is NOT flowing out the top water return flow tube or it is flowing weakly:

First: repeat steps outlined in “Setup and Usage” section (pages 1-4)

If the optional temperature control attachment has been purchased with the system:
Ensure that the temperature control knob is turned to maximum cold (“COLDEST”). Always
adjust the control knob to ensure there is a flow of water from the top water return flow tube.

Check the filter sponge and be sure it is clean and unobstructed. See instructions on page 4.

4. Cleaning and Extended Storage

Note: It is not possible to remove all of the water from the bladder, system and tubing;
therefore before extended storage follow these instructions.

(a)  To clean the system after you have finished all of your cold therapy treatments and prior to

extended storage, fill the reservoir with water to the fill line and mix in 2 tablespoons of
liquid bleach. 

(b)  Connect the bladder to the system and turn the unit on.

(c)  Allow the water and bleach mixture to circulate for 5 minutes. This will help keep the

system clean.

(d)  Turn unit off and unplug the power supply.

(e)  Carefully drain the water and bleach mixture out of the reservoir. Store in a cool dark place

leaving the lid slightly open. (Exposure to sunlight and extreme heat may damage hoses
and unit.)

(f)  When reusing the unit after extended storage, remove the filter sponge from the pump

motor and clean or replace. Then fill the reservoir with clean water, connect the old bladder
and turn the pump on to circulate the water through the hose and bladder and flush the
remaining water mixture out of the system (use the old bladder pad for this procedure,
discard the bladder pad after cleaning is completed). 

A new bladder pad should be

purchased when using the system after extended storage. 

A new filter sponge and

bladder pad may be purchased from Polar Products Inc. Call 800.763.8423 or order on
our website at www.polarproducts.com

(g)  Dispose water and refill the reservoir with fresh ice and water. Follow the operating

instructions in this manual before use.

(h)  Before you use the unit, inspect hoses and new pad for any tears or breaks. Contact Polar

Products Inc. to purchase new therapy bladder pad and filter.

5

TROUBLESHOOTING GUIDE

Содержание Active Ice

Страница 1: ...er with Compression Wraps 8 Cautions and Warnings 9 Warranty and Service 10 Accessories 11 System Components 12 Patients and Medical Professionals should carefully read all of the following instructions and warnings before using and operating the Polar Active Ice Cold Therapy System Polar Products Inc will not be liable for injuries that result from misuse or misapplication of this system ...

Страница 2: ... to 5 or Coldest FILL TO TOP WITH ICE FILL TO THIS LINE WITH WATER DO NOT OVERFILL WITH WATER 2 Use Only as Prescribed Patients first need to be assessed by their physicians for contraindications risk factors and medical history before using this product Supervision may be necessary to ensure precautions are followed especially with young children and the elderly Adhere strictly to your practitione...

Страница 3: ...herapy System includes an electronic digital timer The digital timer may be turned on by pressing the on off button The timer will automatically set at 30 minutes and turn the unit on The default 30 minute setting is a safety feature as many doctors recommend 30 minute therapy sessions followed by skin inspection If your doctor directs you to use the unit for longer periods of time an additional 3...

Страница 4: ...ou experience any adverse reactions such as increased pain burning itching increased swelling blisters increased redness discoloration welts other changes in skin appearance or any other reaction identified by your practitioner 2 Ending the Therapy Session Turn the unit off and disconnect the power supply Press the tabs on the female couplings and gently pull the therapy bladder hose tubes from the...

Страница 5: ...he system after you have finished all of your cold therapy treatments and prior to extended storage fill the reservoir with water to the fill line and mix in 2 tablespoons of liquid bleach b Connect the bladder to the system and turn the unit on c Allow the water and bleach mixture to circulate for 5 minutes This will help keep the system clean d Turn unit off and unplug the power supply e Carefully ...

Страница 6: ...Note the line inside the cooler If the optional temperature control attachment has been purchased with the system Ensure that the temperature control is turned to maximum cold COLDEST If the unit is leaking at the couplings If a leak exists turn off the unit Disconnect and reconnect the couplings make sure both sides are clicked in properly When connecting the couplings ensure that the silver tab ...

Страница 7: ...tional insulation barrier over a sterile dressing that does not completely protect the skin Never use this non sterile universal bladder directly on the skin Position the bladder on the target body part with the clear side toward the body Attach one side of the belt to the bladder then wrap the belt around the bladder and body part Ace Bandage style and then attach the other side of the belt to th...

Страница 8: ... black elastic band secures the wrap to the arm and the long black strap goes around the chest Adjust for comfort KAW LAW SAW CAUTION Do not over tighten the elastic belt s or pinch the tubes or bladder The optional compression wraps and other accessories may be purchased from our website or by calling us 1 800 763 8423 PolarProducts com Using the Universal Bladder with the Optional Compression Wr...

Страница 9: ...rolonged therapy unconscious or incapacitated patients patients with diabetes and those with poor circulation Application areas with compromised local circulation Potential wound healing problems including those from multiple surgical procedures Circulatory syndromes such as Raynaud s disease Buerger s disease peripheral vascular disease vasospastic disorders sickle cell anemia and hypercoagulable...

Страница 10: ...S WARRANTY AND SERVICE Polar Products Inc warrants that the Active Ice System is fit for use under the normal use for which it is intended and free of any defects in materials and or workmanship for 180 days from the date of initial purchase Polar s obligation under this warranty is limited to the replacement or repair of any defective part s of this product If you encounter a problem with your Act...

Страница 11: ...p has a comfortable snug fit and provides effective cold therapy to the head Will work with a standard AIS system Provides cooling relief to the head by continuously circulating cool water throughout the cap Ease the pain of migraines with cold therapy Power take off adapter PTOA works with standard car power outlets and plugs right into the AIS unit PTOA Male CM Female CF Replacement couplings The...

Страница 12: ...nder the direction and supervision of a licensed healthcare practitioner NOTE Due to health and hygiene regulations all Active Ice Cold Therapy products and parts are non returnable POLAR PRODUCTS 3380 Cavalier Trail Stow OH 44224 800 763 8423 polarproducts com Universal Therapy Bladder For additional bladder options please see the bottom of page 7 UB Cold Therapy System COMPONENTS Active Ice Cool...