mpcal=
(Multipoint Calibration)
Description
Initiates a multipoint calibration. This command is only valid when used in conjunction with
Enter Continuous Sampling Mode (
go
). The TCM2.5/2.6 samples its sensors and adds the
data point to the current set of calibration data.
Syntax
mpcal=e<cr>
:<cr><lf>
go<cr>
.......
......
take data in continuous mode
h<cr>
(or
h
)
:<cr><lf>
mpcal=d<cr>
Valid Values
Input
Description
e
e, enable multipoint calibration
d
d, disable multipoint calibration (default setting)
Table 8: Valid
mpcal=
Values
Output
HnVnMn.nn<cr><lf>
The TCM2.5/2.6 only reports a calibration score after it has been put in Standby mode (
h
), and multipoint
calibration has been disabled (mpcal=d)
Example
mpcal=e<cr>
:<cr><lf>
go<cr>
........
.......
take data in continuos mode
h
<
cr>
:<cr><lf>
mpcal=d<cr>
H4V3M5.00:<cr><lf>
(This is an example of a poor score)
Query Response
mpcal?
:mpcal=
Related Commands
lc? (Query Last Calibration Score)