42
P/N 2018 - 10 B Rev A 6 – Release
Immunity Radio Frequency Guideline
The QuickSee is manufactured to work in the electromagnetic environment as specified in the below
table. The user / operator should make sure it is used within these environmental conditions.
Immunity
Test
Test Level
EN 60601-1-2
Conformity
Level
Electromagnetic
Environment Guide
Radiated RF
IEC 61000-4-
3: 2006 +A1:
2007 +A2: 2010
3 V/m from 80MHz to 2.7GHz
3 V/m from
80MHz to
2.7GHz
The portable and mobile
devices with RF commu-
nication should not be
used near any part of the
device, including cables,
except when they follow
the recommended safety
distance for separation,
calculated with the
equation applicable to the
frequency of the transmit-
ter. Recommended safety
distance for separation:
D= 1.2 x √P from 150kHz
to 80MHz, D= 1.2 x √P
from 80MHz to 800MHz,
D= 2.3 x √P from 800MHz
to 2.5GHz. Where P is the
nominal maximum power
of the transmitter output
expressed in Watts (W) as
specified by the transmit-
ter manufacturer and d is
the recommended separa-
tion distance expressed in
meters (m).
Proximity Fields
RF
IEC 61000-4-3
: 2006 +A1:
2007 +A2: 2010
Test
Frequency
(MHz)
Pulse
Modulation
(Hz)
Immunity
Level
(V/m)
Conformity
Level
(V/m)
385
18
27
27
450
FM + 5 Hz
deviation:
1 kHz sine
28
28
710
745
780
217
9
9
810
870
930
18
28
28
1720
1845
1970
217
28
28
2450
217
28
28
5240
5500
5785
217
9
9
Conducted
Disturbances
RF
IEC 61000-4-6:
2013
Frequency range: 150 kHz to 80
MHz
3 V RMS outside the ISM band, 6 V
RMS in the ISM band
Frequency
range: 150 kHz
to 80 MHz
3 V RMS
outside the ISM
band, 6 V RMS
in the ISM band
The field intensity of the fixed RF transmitters, as checked in an electromagnetic investigation of the