Code
Explanation
Comments
H161
Temperature
The temperature of the tube head is
too high.
Wait for a few minutes for the
tube head to cool down.
H162
The temperature of the lift motor is
too high.
Wait for a few minutes for the
lift motor to cool down.
H165
The temperature of the tube head is
too high for the selected exposure
values.
Wait for a few minutes for the
tube head to cool down.
H166
The maximum tube head energy
level was exceeded.
Wait for a few minutes for the
tube head to cool down or
use lower exposure values.
H170
User related
messages
Wrong licence code.
Check the licence code.
H171
Timeout in production test.
H181
The imaging process was cancelled
in Planmeca Romexis.
H182
Timeout in image data transmission. Exposure was interrupted.
Contact your service
technician for help.
H186
No IP address defined for 3D
sensor.
H187
Problem during image data
transmission.
Exposure was interrupted.
Contact your service
technician for help.
H189
The screen was touched during
exposure.
Exposure was interrupted.
H190
Protouch-CPU communication
failure.
H191
3D sensor communication failure.
H192
Workstation communication failure.
H193
Invalid scan settings.
H194
CPU connection not established.
H195
Request timed out while waiting for
CPU to respond.
H196
Version mismatch in communication
interfaces.
H197
Workstation communication
disabled.
H199
Video streaming failed.
13 Help messages
68 Planmeca Viso G7
User's manual