NOTE
EMC requirements have to be considered, and the equipment must be
installed and put into service according to the specific EMC information
provided in the accompanying documents.
NOTE
Portable and mobile RF communications equipment can affect the X-ray
unit.
NOTE
External equipment intended for connection to signal input, signal output or
other connectors, shall comply with relevant IEC standard (e.g. IEC 60950
for IT equipment and the IEC 60601 series for medical electrical equipment).
In addition, all such combinations - systems - shall comply with the standard
IEC 60601-1, Safety requirements for medical electrical systems. Equipment
not complying to IEC 60601 shall be kept outside the patient area (more
than 2m (79 in.) from the X-ray unit). Any person who connects external
equipment to signal input, signal output or other connectors has formed a
system and is therefore responsible for the system to comply with the
requirements of IEC 60601-1. If in doubt, contact your service technician or
local representative for help.
NOTE
Contact your service technician if you notice a decrease in image quality.
NOTE
If you take an exposure but the image does not appear in the Planmeca
Romexis program, you can import the image manually into Romexis. Refer
to the Planmeca Romexis User's Manual for details.
NOTE
Do not handle liquids near or on the X-ray unit.
NOTE
Never place or hang any objects on any part of the X-ray unit.
NOTE
Make sure that neither you nor your patient can get caught or hooked up on
any part of the X-ray unit. Keep loose items of clothing, hair and jewellery
tucked away safely.
NOTE
Do not touch the arm structures when the X-ray unit is moving.
5 Safety precautions
User's manual
Planmeca Viso G7 9