Cables
ControlBox to PoE
RJ45 10m OR 15m
PoE to LAN
RJ45 10m OR 15m
PoE power supply
Phihong Single Port Injector
Type
PSA16U-480 (POE)
POE15M-1AF/
POE15M-1AFE-R
Input voltage
100-240 VAC (50-60 Hz)
100-240 VAC (50-60 Hz)
Output voltage
48 VDC
56 VDC
Max. output current
0.32 A
0.275 A
Insulation voltage primary-
secondary
3000 VAC
3000 VAC
13.3 USB ControlBox
Dimensions
112 x 46 x 24 mm (4.41 x 1.81 x 0.94 in.)
Cables
Fixed USB 2.0 power supply cable 2 or 5m (6.6
or 16.4 ft)
Power input
2.5 W
13.4 Supported operating systems
•
Windows OS (64 bit)
•
Mac OS X
NOTE
Check the latest up-to-date system recommendation from the following
http://www.planmeca.com/System-requirements
13.5 Operating environment
Planmeca ProSensor is for indoor use only. The equipment is installed on
the wall or on/under the table. The user moves the sensor into the operation
position by hand.
The room and operation must comply with the x-ray safety shielding
requirements according to radiation safety regulation in the country.
The system is used by dental care professionals.
Prior to installation of the system check that the local conditions are
compatible with the appliance design.
The temperature of the operating environment should be b 15°C
and + 40°C.
The relative humidity of the operating environment should not exceed 60%.
Atmospheric pressure range should be between 700 hPa - 1060 hPa.
13.6 Transportation and storage environment
Transportation and storage temperature -10°C - +60°C.
The relative humidity during transportation and storage should not exceed
95%.
13 Technical specifications
User's manual
Planmeca ProSensor® HD 21
Содержание ProSensor HD