Wear protective gloves.
1.
Place the image plate in the light protection cover into the patient's
mouth.
When doing this, make sure that the active side of the image plate points
towards the X-ray tube.
2. Set the exposure time and setting values on the X-ray unit (see section
"X-ray unit settings" on page 30).
3. Record an X-ray image.
The image data must be scanned within 30 minutes.
Preparing for scanning
CAUTION
Light erases the image data on the image plate
Never handle exposed image plates without the light protection cover.
Wear protective gloves.
1. Remove the image plate with the light protection cover from the patient's
mouth.
WARNING
Contamination of the unit
Clean and disinfect the light protection cover before removing the image
plate.
2. In the event of heavy soiling, e.g. from blood, dry clean the light
protection cover and protective gloves, e.g. wipe with a clean cellulose
cloth.
4 Usage
User's manual
Planmeca ProScanner 2.0 41
Содержание ProScanner 2.0