38
Maintenance
5
Maintenance
5.1
Calibration
The data logger and sensor should be calibrated annually by an authorised service
provider.
5.2
Cleaning the SL3
1. Gently wipe with a damp cloth and clean water.
2. Wait 2 minutes.
3. Allow to dry.
5.3
Disinfecting the SL3
·
Note that disinfectants are potentially hazardous to health. Ensure that you
obtain a material safety data sheet (MSDS) before use and follow the
instructions contained therein.
·
Only cleaning and disinfecting materials compatible with the equipment
must be used.
1. Gently wipe with a damp cloth and sterile water.
2. Wait 2 minutes.
3. Repeat steps 2 and 3, three more times.
4. Clean surfaces with a 70 % isopropyl alcohol wipe.
5. Wait 15 minutes.
6. Gently wipe with a damp cloth and sterile water
7. Repeat steps 4 to 6.
8. Allow to dry.