![Planer kryo connect Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/planer/kryo-connect/kryo-connect_getting-started_1562048016.webp)
Getting Started
14
RS232 / RS485 Connections
The
kryo-connect
is provided with a 9 way D-Type plug that can be
configured for both RS232 and RS485 interfaces. The actual configuration is
determined by dip-switches inside the unit and is normally factory
configured. The information below can be used if the configuration needs to
be changed or if interconnecting cables need to be made.
It is important that the kryo-connect is connected to ground to
prevent charge build up. On RS232 connections this is achieved
through the cable screen and signal ground. On RS485
connections, if no ground is provided through the cable, the
chassis ground
should be connected to a suitable ground
reference.
Pin outs
Port type
DIP switch *
1
2
3
4
5
6
7
8
9
RS232 –
Wired as
DTE.
Requires null
modem cable
to connect to a
PC
-
R
X
T
X
-
G
N
D
-
R
T
S
C
T
S
-
RS485 full
duplex
tx
-
rx
-
tx
+
rx
+
RS485 Half
duplex
A
-
B
+
RS485
terminate RX
120 Ohms
Connects a 120 Ohm resistor across the RX
pair
RS485
terminate TX
120 Ohms
Connects a 120 Ohm resistor across the TX
pair
* grey switches means the setting does not matter or is determined by another
requirement such as termination
Содержание kryo connect