52
Additional information
5.3.8
Monitoring
Feature
Controller
Local area network (LAN)
10 Base T Ethernet - RJ45 shielded. Modbus-TCP-IP
protocol.
Independent temperature
monitoring
Independent sensors can be fitted to the monitoring
ports; see
Side view
section.
Recommended sensor type: PT100 Class A to
EN60751.
Maximum diameter: 2.51 mm.
Contact your service provider for more details and available options.
5.3.9
Fuses
Warning
To avoid risk of fire, fuses must always be replaced with the same type and rating.
o
Fuses should only be replaced by suitably trained service personnel.
o
Fuses should only be replaced after the cause of the original failure has been
determined and corrected as appropriate.
Fuse
Location
Type
F1, F2
Mains inlet
T 3.15A L 250V 5 x 20 mm
12