Routine maintenance and troubleshooting
4
Routine maintenance and troubleshooting
4.1
Regular checks
Daily
Check bubbles can be seen through the liquid level indicator.
See
Checking the liquid level indicator
.
If there is insufficient
water to cover the dip tube in the bottle, replace the humidifier.
Check the humidifier tubing to ensure there is no build up of
condensation. If condensation is forming in the tubes, refer to the
Condensation
section.
When
samples are
added or
removed.
Check the humidifier tubing to ensure there is no build up of
condensation. If condensation is forming in the tubes, refer to the
Condensation
section.
Every 4
months
Check the battery. See
Checking the battery
.
Annually
Calibrate and service the BT37-02. See
Calibration and servicing
.
4.2
General cleaning
Warning
Bleaches are corrosive and may damage sensitive components and metal surfaces
within the chamber.
Switch off the BT37-02 and disconnect the mains supply before cleaning. See
Switching off
.
Always allow the unit to dry fully before reconnecting the mains supply.
Note that disinfectants are potentially hazardous to health. Ensure that you obtain a
material safety data sheet (MSDS) before use and follow the instructions contained
therein.
Caution
The person responsible for the equipment must ensure that:
o
the unit is decontaminated if hazardous material is split onto or into the
equipment.
o
only cleaning and disinfecting materials compatible with the equipment are used.
Incompatible materials may cause a hazard by reacting with the equipment or
materials contained within.
These instructions are for the exterior of the device only.
36
43
43
36
37
30
Содержание BT37-02
Страница 1: ...INSTRUCTIONS FOR USE BT37 v2 MA200214 M EN 2 0 24...
Страница 2: ......
Страница 5: ...Introduction...
Страница 17: ...Installation...
Страница 21: ...Operation...
Страница 31: ...31 Operation 4 You can now switch off the mains display and disconnect the power cord from the mains inlet...
Страница 32: ......
Страница 33: ...Routine maintenance and troubleshooting...
Страница 46: ......
Страница 47: ...Additional information...
Страница 55: ...55 Index view side 12 W warnings 8 warranty 6...
Страница 56: ......