vi | Dome EX Display
Safety precautions
External equipment intended for connection to signal input, signal
output, or other connectors, must comply with the relevant IEC standard
(EN/IEC 60601-1 series for medical electrical equipment). In addition, all
such combinations (systems) must comply with the standard IEC 60601-1-1,
Safety requirements for medical electrical systems.
Equipment not complying to IEC 60601 must be kept outside the patient
environment, as defined in the standard as at least 1.5 meters from the
patient or the patient support.
Any person who connects external equipment to signal input, signal output,
or other connectors has formed a system and is therefore responsible for the
system to comply with the requirements of IEC 60601-1-1. If in doubt, speak
with a qualified technician.
GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN EQUIPMENT IS
CONNECTED TO AN EQUIVALENT RECEPTABLE MARKED “HOSPITAL ONLY” OR
“HOSPITAL GRADE.”
Safety tips
• Never open the display case, even when the power is off. Dangerous
voltage inside may cause electric shock or death.
• To avoid damage to the display, use the grounded power supply and
video cable supplied by Planar, or use certified replacements.
• Be sure the display is electrically grounded. You must connect the third
grounding pin on the US power cord to a grounded outlet. The European
power cord does not have a third grounding pin, but it must be plugged
into a grounded outlet.
• If you cannot insert the plug into the outlet you plan to use, have a
licensed electrician replace the outlet with a properly grounded outlet.
If the power cord connects directly into the computer, make sure the
computer is grounded.
• Keep the display dry if it is part of a surgical system. The display lacks
protection against liquids or spills.
• In locations where 240V outlets are used, connect the Dome EX
display only to a center-tapped, 240V, single-phase supply (only for
Canada and the United States).
Содержание Dome EX
Страница 1: ...Dome EX Display User s Guide ...