Error messages

Error 1: Echo Bad

Very turbulent flow, or too high air content in the medium; no measurement possible.
The reason may be that the medium generally has a high proportion of air, or the propor-
tionate air
is dispensed by turbulence in the measuring system. See point 6 (Selection of the installa-
tion location).

Error 2: No Fluid

No media in measuring pipe. No measurement is possible. Fill system with medium.

The DU06 indicates no negative flow. If there is 0 flow indication while there is flow, flow
direction has to be changed in the program. 

Functional check

It is the responsibility of the operator to test the function periodically and to check the de -
vice visually. periodically perform a functional test or a visual inspection.
A functional test of the contacts can be carried out in the installed or removed state, with
manual displacement of the float or by filling the container. It should be noted that no func-
tional processes are triggered by the functional test.

DU06 Instruction manual 09/2019

page  11

Содержание DU06 Series

Страница 1: ...Instruction Manual DU06 Ultrasonic flowmeter with display PKP Prozessmesstechnik GmbH Borsigstra e 24 D 65205 Wiesbaden Nordenstadt Tel 49 0 6122 7055 0 Fax 49 0 6122 7055 50 Email info pkp de...

Страница 2: ...vent of damage due to improper use non observance of this operating manual use of insufficiently qualified personnel and unauthorized modification of the device Proper Usage The ultrasonic flow meter...

Страница 3: ...on versions are available version made of brass or stainless steel flow max 80 l min 1 version made of stainless steel flow max 180 l min 2 version made of stainless steel flow max 1000 l min All vers...

Страница 4: ...urrent voltage frequency output 4 pole plug included in delivery Pin 1 4 20mA 0 10V Freq Out Pin 2 GND Pin 3 GND Pin 4 24VDC 15 Relay outputs 6 pol plug included in delivery Pin 1 Relay 2 normally ope...

Страница 5: ...parameter Output Parms Menu 1 Relay 1 setting switch point of relay 1 Relay 2 setting switch point of relay 2 If the actual measured flow exceeds falls below one of the adjusted switch points the re...

Страница 6: ...erated Freq Out 0 Hz is generated Assignment of analog output signal If the analogue output is used the following assignment must be observed 20 mA corresponds to the Max Flow Limit 4 mA corresponds t...

Страница 7: ...c Litre per second l min Litre per minute l h Litre per hour m min Cubic meter per minute m h Cubic meter per hour gal sec Gallons per second gal min Gallons per minute gal h Gallons per hour Percent...

Страница 8: ...e flow passing the DU06 during power failure was not counted Parameter protection Pin Pin Code is for protecting parameter of the device Changing them without authority will be avoided If the Pin is a...

Страница 9: ...fset This function is used for adaption of the characteristic diagram up to formula Qkorr Qist DynOffset Qist QmaxPhys With Qist measured flow Qkorr corrected flow QmaxPhys max flow in pipe at v 5m se...

Страница 10: ...s and protected due to manufacturers code 4 20 mA K value of current output 4 20 mA K Value This function is for special applications and protected due to manufacturers code 0 10 V K value of voltage...

Страница 11: ...dium The DU06 indicates no negative flow If there is 0 flow indication while there is flow flow direction has to be changed in the program Functional check It is the responsibility of the operator to...

Страница 12: ...time difference is proportional to the flow rate In contrast to other types of ultrasonic flowmeters the sensors are oriented parallel to the flow stream This results in an extremely compact meter wh...

Страница 13: ...Dimensions Order Code Order number DU06 Ultrasonic flowmeter and counter with local indication E V R025 0 0 Material version M brass housing sensors st steel with R male connection only E st steel ho...