Instrument test function <Instr. Test> Menu 4
Indication of supply voltage - Supply Voltage
The device works like a Voltmeter and indicates the actual supply voltage.
Puss button <P> to finish the measurement.
Advice!
The allowed voltage is: 24 VDC +/-15 %, max. Ripple 100 mV
If low-grade power supply units are used, restricted function of the DU06 respectively inac-
curate
Measurement can appear. It is forbidden to operate inductive load with the same power
supply than the DU06
4-20 mA offset of analog output
-
4-20 mA Offset
This function is for special applications and protected due to manufacturers code.
4-20 mA K-value of current output - 4-20 mA K-Value
This function is for special applications and protected due to manufacturers code.
0-10 V K-value of voltage output- 0-10 V K-Value
This function is for special applications and protected due to manufacturers code.
Flow test function
-
Flow Test
This function simulates flow rates between zero and
Max Flow Limit,
for demonstration
and test purposes
during configuration of the plant
.
The display, the relays contacts as
well as 4-20 mA, 0-10 V, Fout and the interface is operated. Thereby the DU06 works
without pipe and media.
Relays Test Function –
Relays Test
Relays test function operates the relays alternating (relays on/off) 5.