The
alarm module
can in addition to the
above power module, provide 3 customer
individually settable alarms and 2 inputs. The
module is supplied with a 2 meter fly-lead
cable.
Each of the 3 alarm contacts can be
configured as high, low, window and totalizer
alarm. Once the alarm becomes active, the 3
potential free alarm contacts can be reset in
many different ways. Automatic, through the
touch screen or with one of the 2 input
signals. Other options like hysteresis settings
and delays are also available.
For detailed information see
4.11.2 Alarm
Menu
in this manual.
Both the power and the alarm module are
mounted similar to the battery module (see
3.16 How to replace the battery
). They
replace the battery module, so you cannot
use these modules in combination with
battery power.
Generic specifications optional alarm module
Power input:
Voltage supply range : 12 - 30 VDC (Power supply hereafter refered to as the 24 Vdc power
supply)
Max. current : 300 mA
Protection input: Polyfuse (trip current > 500mA) and reverse polarity protected
Switch ratings switch 1 + 2 (SPST hard contact):
Maximum current
0.5 A
Maximum voltage
30 Vdc
Switch ratings switch 3 (DPST contact):
Maximum current
1 A
Maximum voltage
30 Vdc
Inputs 1 + 2 (Opto-couplers):
Voltage range (polarity sensitive): 5 – 30 Vdc (@ 5mA max)
Min. recommended pulse width
100 msec (Sample interval: 20 msec)
2.16. Calibration
Each measuring device is supplied with a factory calibration report. The calibration is traceable to
American and/or European standards. On request we can also provide DAkkS calibration (German
Accreditation Body).
Figure 5: Power and Alarm module wiring