PIE Evolution piecal 211 Скачать руководство пользователя страница 9

Page 9

Troubleshooting RTD Sensors

When troubleshooting a problem with an RTD input it is useful to check that the sensor and the 
wiring to the instrument is operating properly.

The PIE 211 automatically detects 2, 3 and 4 wire RTD connections with a patented circuit. It will 
also display the connections on the display and indicate when there is a missing connection due to 
a loose connector, corrosion or a broken wire.

Here is an example of the PIECAL 211 reading a 
sensor with all 4 wire connected.

Here is an example where connections are made to a 
4 wire sensor and the 211 indicates that only Wires 1, 
2 & 4 are connected. There may be a loose connection 
or a break in wire 3 somewhere between the sensor 
and the 211.

Troubleshooting RTD Sensors

This is much simpler and quicker than going through 
the process of testing each pair of wires to figure out 
which, if any, connection is loose or which wire is 
broken.

Содержание Evolution piecal 211

Страница 1: ...itch Calibrate quickly with automatic output stepping Choose between 2 3 5 11 and 21 steps to automatically increment the output in 100 50 25 10 or 5 of span Select the step time to match your system...

Страница 2: ...to start the automatic stepping READ Press and hold to transfer the current temperature into the EZ Check HI LO points This clears the HI LO temperature readings which will update as the temperature...

Страница 3: ...cycle through the various RTD Pt Cu Ni at different base resistances and alpha values DISPLAY OHMS If DISPLAY OHMS is ON the resistance associated with the RTD temperature will appear in small digits...

Страница 4: ...d LO EZ CHECK in 25 increments 0 25 50 75 100 11 steps between the HI and LO EZ CHECK in 10 increments 0 10 20 80 90 100 21 steps between the HI and LO EZ CHECK in 5 increments 0 5 10 90 95 100 Note A...

Страница 5: ...LO position and turn the e EZ Dial knob until the desired temperature is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 3 Instantly ou...

Страница 6: ...alibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated READ Plug in the 2 3 or 4 wires from the sensor and...

Страница 7: ...the knob for faster dialing with 10 or 1 00 10 0 ohm increments The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator Calibrating RTD Instruments 3 2 2 1 THE...

Страница 8: ...splay with the current temperature reading Move the EZ CHECK switch to MAX to see the highest temperature reading and to MIN to see the lowest temperature reading Press and hold the knob e to clear th...

Страница 9: ...connection due to a loose connector corrosion or a broken wire Here is an example of the PIECAL 211 reading a sensor with all 4 wire connected Here is an example where connections are made to a 4 wir...

Страница 10: ...460 0 to 720 0 720 0 to 850 0 0 2 0 3 0 4 0 5 328 0 to 338 0 338 0 to 860 0 866 0 to 1328 0 1328 0 to 1562 0 0 4 0 6 0 8 0 9 Pt 1000 Ohm DIN IEC JIS 1989 1 3850 0 00385 200 0 to 230 0 230 0 to 550 0 5...

Страница 11: ...onds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Excitation Current 0 9 mA to 401 Ohms 0 4 mA to 2010 Ohms nominal Normal Mode Rejection 50 60 H...

Страница 12: ...ipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No responsibility is accepted for damage...