EZ-CHECK™ SWITCH

SOURCE:

 Instantly output two preset RTD 

temperatures by moving the EZ-CHECK™ switch to 
the 

“LO” 

position or 

“HI”

 position. For fast three 

point checks select the 

“DIAL”

 position. The 

PIECAL 211 will remember the last 

“DIAL” 

value, even with the power off.
These values can easily be changed to suit the 
calibration requirements. The temperatures stored in 
the HI and LO positions are also used for Auto 
Stepping.

READ:

 Slide the switch to the DIAL position. The 

PIECAL 211 will display the current temperature from 
the RTD sensor.  Slide the switch to HI and the 
highest temperature measured since turn-on or reset 
will be displayed;  slide the switch to  LO and the 
lowest temperature measured since turn-on or reset 
will be displayed.

w

 

SOURCE/OFF/READ Switch

Select “

SOURCE

” to output in °C, °F or ohms.  

Select “

READ

” to read an RTD sensor or ohms.

e

 

EZ-DIAL™ KNOB

SOURCE:

 Turn the knob to adjust the output 

level. Turn clockwise to increase the output, 
counter clockwise to decrease the output in 0.1°  
steps at a time. Push down and turn the 
EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store 
desired EZ-Check™ HI/LO points in SOURCE 
mode. 

Continue to press and hold the knob for 

two more seconds to start the automatic 
stepping.

READ: 

 Press and hold to transfer the current 

temperature into the EZ-Check™ HI/LO points. 
This clears the HI/LO temperature readings 
which will update as the temperature changes.

Double click the knob to get into the PIECAL 211 
menu. Use the menu to  select RTD or Ohms, °C 
or °F, 400Ω or 2000Ω, RTD Type, Backlight On/
Off, Step Size, Step Time and Auto Off On/Off.

CHANGING BATTERIES

Low battery is indicated by a battery symbol on 
the display. Approximately one to four hours of 
typical operation remain before the PIECAL 211 
will automatically turn off. To change the batteries; 
remove the rubber boot, remove the battery door 
from the back of the unit by sliding the door 
downward. This allows access to the battery 
compartment. Replace with four (4) “AA” 1.5V 
batteries being careful to check the polarity. 
Replace the battery door and replace the boot. All 
stored configuration options (RTD Type, 
EZ-CHECK  Memories, etc., are reset to factory 
settings when the batteries are removed.

Note:

 Alkaline batteries are supplied and 

recommended for maximum battery life and 
performance.

Basic Operation

3 2 2 - 1

THERMOCOUPLE

CALIBRATOR

J, T, E, K, mV

HI

Pt100

 

α

=385

AUTOMATED

RTD

CALIBRATOR

READ

SOURCE

HI

SET

LO

MAX

READ

MIN

OVERLOAD

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

OFF

°C

          

3      2           1     4

100.0

+

-

-

+

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

 TestEquipmentDepot.com 

Содержание Evolution piecal 211

Страница 1: ...itch Calibrate quickly with automatic output stepping Choose between 2 3 5 11 and 21 steps to automatically increment the output in 100 50 25 10 or 5 of span Select the step time to match your system...

Страница 2: ...to start the automatic stepping READ Press and hold to transfer the current temperature into the EZ Check HI LO points This clears the HI LO temperature readings which will update as the temperature...

Страница 3: ...cycle through the various RTD Pt Cu Ni at different base resistances and alpha values DISPLAY OHMS If DISPLAY OHMS is ON the resistance associated with the RTD temperature will appear in small digits...

Страница 4: ...d LO EZ CHECK in 25 increments 0 25 50 75 100 11 steps between the HI and LO EZ CHECK in 10 increments 0 10 20 80 90 100 21 steps between the HI and LO EZ CHECK in 5 increments 0 5 10 90 95 100 Note A...

Страница 5: ...LO position and turn the e EZ Dial knob until the desired temperature is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 3 Instantly ou...

Страница 6: ...alibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated READ Plug in the 2 3 or 4 wires from the sensor and...

Страница 7: ...the knob for faster dialing with 10 or 1 00 10 0 ohm increments The OVERLOAD indicator will light if excessive voltage or current is detected by the calibrator Calibrating RTD Instruments 3 2 2 1 THE...

Страница 8: ...splay with the current temperature reading Move the EZ CHECK switch to MAX to see the highest temperature reading and to MIN to see the lowest temperature reading Press and hold the knob e to clear th...

Страница 9: ...connection due to a loose connector corrosion or a broken wire Here is an example of the PIECAL 211 reading a sensor with all 4 wire connected Here is an example where connections are made to a 4 wir...

Страница 10: ...460 0 to 720 0 720 0 to 850 0 0 2 0 3 0 4 0 5 328 0 to 338 0 338 0 to 860 0 866 0 to 1328 0 1328 0 to 1562 0 0 4 0 6 0 8 0 9 Pt 1000 Ohm DIN IEC JIS 1989 1 3850 0 00385 200 0 to 230 0 230 0 to 550 0 5...

Страница 11: ...onds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Excitation Current 0 9 mA to 401 Ohms 0 4 mA to 2010 Ohms nominal Normal Mode Rejection 50 60 H...

Страница 12: ...ipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No responsibility is accepted for damage...

Отзывы: