Product information
7
Copyright © 2007 Pico Technology Limited. All rights reserved.
ps3425.en
2.3
Connector diagrams
Front view (3425)
Rear view
1. Ch A. Input channel A. As this is a
the channels have
non-standard input impedances and cannot be used with passive attenuated
scope probes such as the x1/x10 probes supplied with conventional oscilloscopes.
2. Ch B. Input channel B. Has the same characteristics as Ch A.
3. Ch C. Input channel C. Has the same characteristics as Ch A.
4. Ch D. Input channel D. Has the same characteristics as Ch A.
5. Ground. Can be used with the current clamp adaptor supplied to convert one or
more of the differential inputs to single-ended inputs. Can also be used to ground
the screen of the shielded twisted-pair cable supplied.
The scope's ground is connected to the PC's ground through the USB
cable. You MUST NOT rely on the scope's ground as a protective safety
ground.
6. LED. Lights up when the oscilloscope is first powered up, switches off when the
PicoScope software begins to run, and then flashes while the oscilloscope is
capturing data.
7. USB port. Compatible with
1.1 and USB 2.0 ports.
10