pickering

SECTION 6 - TROUBLE SHOOTING

REED RELAY MODULE 40-110/115/120/125

Page 6.1

pickering

SECTION 6 - TROUBLE SHOOTING

INSTALLATION PROBLEMS

The Plug & Play functionality of Pickering switch cards generally ensures trouble-free installation.
If you do experience any installation problems you should first ensure that all cards are properly seated in their 

slots. Improperly mated cards may go undetected by the operating system, or may be detected as a card of an 

unknown type. They can also cause the computer to freeze at various stages in the boot sequence.
If your system employs a remote control interface you should check the integrity of all remote control interface 

links. When the system is powered up, and during Windows start-up, you should expect to see periodic activity 

on the remote control interface RX/TX (yellow) indicators, clearing to leave only the PWR/LNK (green) LEDs 

illuminated. The RX/TX indicators should show activity when you attempt to access a card.

DIAGNOSTIC UTILITY

The Pickering Diagnostic Utility (accessible through the Programs>>Pickering>>PXI Utilities menu) generates 

a diagnostic report of the system’s PCI configuration, highlighting any potential configuration problems. Specific 

details of all installed Pickering switch cards are included. All the installed Pickering switch cards should be listed 

in the “Pilpxi information” section - if one or more cards is missing it may be possible to determine the reason by 

referring to the PCI configuration dump contained in the report, but interpretation of this information is far from 

straightforward, and the best course is to contact Pickering support: 

[email protected]

, if possible 

including a copy of the diagnostic report.
In  the  “VISA  information”  section,  if  VISA  is  not  installed  its  absence  will  be  reported. This  does  not  affect 

operation using the Direct I/O driver, and is not a problem unless you wish to use VISA. VISA is a component of 

National Instruments LabWindows/CVI and LabVIEW, or is available as a standalone environment.
If VISA is present and is of a sufficiently recent version, the section “Pipx40 information” should present a listing 

similar to “Pilpxi information”.
Please note that the Diagnostic Utility cannot access cards if they are currently opened by some other application, 

such as the Test Panels or Terminal Monitor.

SECTION 6 - TROUBLE SHOOTING

!

Refer to the Warnings and Cautions at the front of this manual

Содержание PXI 40-785

Страница 1: ...underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos I...

Страница 2: ...Page 1 MICROWAVE MULTIPLEXER MODULE 40 785 pickering Issue 3 2 August 2019 User Manual PXI Microwave Multiplexer Module Model No 40 785 pickeringtest com pickering...

Страница 3: ...CKERING INTERFACES ALL RIGHTS RESERVED No part of this publication may be reproduced transmitted transcribed translated or stored in any form or by any means without the written permission of Pickerin...

Страница 4: ...SO 14001 Pickering Interfaces strives to fulfil all relevant environmental laws and regulations and reduce wastes and releases to the environment Pickering Interfaces aims to design and operate produc...

Страница 5: ...hazards arising from dangerous voltages WARNING CAUTION An appropriate safety instruction should be followed or caution to a potential hazard exists Refer to the relevant instructions detailed within...

Страница 6: ...1 1 Section 2 Technical Description 2 1 Functional Description 2 1 Section 3 Installation 3 1 Pre Operation Checks 3 1 Hardware Installation 3 2 Software Installation 3 2 Testing Operation 3 3 Sectio...

Страница 7: ...Page VI MICROWAVE MULTIPLEXER MODULE 40 785 pickering THIS PAGE INTENTIONALLY BLANK...

Страница 8: ...host chassis must be correctly installed and be connected to mains or factory safety ground please refer to your chassis installation manual Diagram Showing Separation Between The Mains Circuit and Se...

Страница 9: ...itably qualified trained users should ensure that the accompanying documentation is fully read and understood before attempting to install or operate the product SAFETY INSTRUCTIONS SAFETY INSTRUCTION...

Страница 10: ...e connectors for 50 versions The 40 785 range of PXI microwave multiplexers are suitable for switching 50 signals up to 40GHz They are available in single or dual 6 channel configurations with relays...

Страница 11: ...nsport Conditions Storage Temperature Humidity Altitude 20 C to 75 C Up to 90 non condensing 15000m PXI CompactPCI Compliance The module is compliant with the PXI Specification 2 2 Local Bus Trigger B...

Страница 12: ...Versions 50 Single 6 Chan Panelmount 50 SMA 2 9 Dual 6 Chan Panelmount 50 SMA 2 9 40 785 541 40 785 542 Single 6 Chan Panelmt 50 SMA 2 9 Term Dual 6 Chan Panelmt 50 SMA 2 9 Term 40 785 541 T 40 785 54...

Страница 13: ...on Legacy and Hybrid Peripheral slots in a 3U PXI Express PXIe chassis Pickering Interfaces LXI or LXI USB Modular Chassis Chassis Selection Guide Standard PXI or hybrid PXIe Chassis from any Vendor M...

Страница 14: ...go to pickeringtest com software Signal Routing Software Our signal routing software Switch Path Manager automatically selects and energizes switch paths through Pickering switching systems Signal rou...

Страница 15: ...Page 1 6 MICROWAVE MULTIPLEXER MODULE 40 785 pickering SECTION 1 TECHNICAL SPECIFICATION THIS PAGE INTENTIONALLY BLANK...

Страница 16: ...wing CPCI Ejector Handle Three four or six slot width Front Panel single slot for the remote E option One or two Front Panel mounted Microwave Multiplexers remotely located via cable with the E option...

Страница 17: ...vides mounting for the microwave relays the remote E option controls the relays via a 1 5m screened cable The relays are energised via control signals from relay drivers U10 and U11 The relay drivers...

Страница 18: ...elative to any other equipment the module s will connect with Ensure the chassis is not connected to the electrical supply 3 Ensure that the designated area for the chassis containing the module is of...

Страница 19: ...r to page VII Ensure that there is adequate ventilation The module should be installed in accordance with the following procedure 1 Ensure that the system is turned OFF but still connected to mains so...

Страница 20: ...t Panel GSFP using the desktop icon or via the Start menu Start Menu Pickering Interfaces Ltd General Soft Front Panel A selector panel will appear listing all installed Pickering PCI PXI or LXI switc...

Страница 21: ...Page 3 4 MICROWAVE MULTIPLEXER MODULE 40 785 pickering SECTION 3 INSTALLATION THIS PAGE INTENTIONALLY BLANK...

Страница 22: ...on on the installation and use of drivers and the programming of Pickering s products in various software environments please refer to the Software User Manual This is available as a download from htt...

Страница 23: ...Page 4 2 MICROWAVE MULTIPLEXER MODULE 40 785 pickering SECTION 4 PROGRAMMING GUIDE THIS PAGE INTENTIONALLY BLANK...

Страница 24: ...Only the correct SMA connector series should be mated with these modules The 40 785 range uses SMA 3 5 SMA 2 9 SMA 1 8 connectors for 40 50 65GHz respectively and the user should only mate the correct...

Страница 25: ...5 CONNECTOR INFORMATION Figure 5 2 Microwave Multiplexer Module 40 785 Connector Positions 40 785 521 751 40 785 522 752 40 785 521 T 40 785 522 T pickering 1 2 3 6 C 4 5 pickering 40 785 Remote Versi...

Страница 26: ...u generates a diagnostic report of the system s PCI configuration highlighting any potential configuration problems Specific details of all installed Pickering switch cards are included All the instal...

Страница 27: ...MICROWAVE MULTIPLEXER MODULE 40 785 Page 6 2 pickering SECTION 6 TROUBLE SHOOTING THIS PAGE INTENTIONALLY BLANK...

Страница 28: ...I modules which are supported in one of Pickering Interfaces Modular LXI Chassis such as the 60 102B and 60 103B no module software update is required If the module was introduced after the LXI chassi...

Страница 29: ...yout U1 U2 U3 U4 U5 U6 U7 R1 R2 R5 R6 R7 R8 R9 R10 R11 R12 R13 R14 C1 C2 J1 C3 C4 C5 C6 C7 C8 C9 C10 C26 C11 C13 C14 C12 C15 C16 pickering C23 R15 R4 U8 U9 U10 U11 HD1 HD2 HD3 HD4 HD5 HD6 HD7 HD8 HD9...

Страница 30: ...785 pickering SECTION 7 MAINTENANCE INFORMATION Figure 7 2 Dual 6 Channel Multiplexer in Panel Mount Format Figure 7 3 Triple 6 Channel Multiplexer in Remote Mount Format Multiplexers are connected to...

Страница 31: ...Page 7 4 MICROWAVE MULTIPLEXER MODULE 40 785 pickering SECTION 7 MAINTENANCE INFORMATION THIS PAGE INTENTIONALLY BLANK...

Страница 32: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...