![Pickering PXI 40-412 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/pickering/pxi-40-412/pxi-40-412_user-manual_1554975024.webp)
Page 4.4
32-CHANNEL DIGITAL I/O MODULE 40-412
pickering
SECTION 4 - PROGRAMMING GUIDE
Input Sub-Units
Model -001 only
Description
Number
Type
Size
1
INPUT
2
ets level of selected in ut channel its
00 = below threshold 2, below threshold 1
01 = below threshold 2, above threshold 1
10 = above threshold 2, below threshold 1
11 = above threshold 2, above threshold 1
2
INPUT
64
ets levels of all in ut channels its each as a ove
Applicable Functions
VISA Driver
Direct Driver
pipx40_readInputPattern
PIL_ReadSub
NOTE: each input channel from 1 to 32 is sampled sequentially.
The precise rate of sampling is undefined.
Input Sub-Unit
Model -111 only
Description
Number
Type
Size
1
INPUT
64
Gets levels of all 32 input channels, 2 bits each, as follows:
00 = below threshold 2, below threshold 1
01 = below threshold 2, above threshold 1
10 = above threshold 2, below threshold 1
11 = above threshold 2, above threshold 1
Applicable Functions
VISA Driver
Direct Driver
pipx40_readInputPattern
PIL_ReadSub