User Manual
P736T0002, applies to P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2021-09-20
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Page 16 / 27
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
Mounting Holders for Petri Dishes, Microscope Slides and
Accessories
Figure 7: M2 threaded holes in the sample holder mounting plane (p. 9) of the motion platform
Requirements
You have read and understood the safety precautions (p. 6).
The P-736.ZRx is
not
connected to the controller.
Tools and accessories
P-545.PD3, P-545.SH3 or P-545.PP3 holder (p. 10)
Four M2 screws of suitable length (included with the P-545.xx3 holder)
Suitable screwdriver
Mounting a holder
1.
Align the holder so that the holding clamps (P-545.PD3 and P-545.SH3) are pointing
upwards or the larger surface of the base body (P-545.PP3) is pointing upwards. Refer
to "Dimensions" (p. 23ff.) for details.
2.
Place the holder into the aperture of the P-736.ZRx from above, so that the mounting
holes in the holder and the sample holder mounting plane (see Figure
P-736.ZRx are in line.
3.
Mount the holder with the M2 screws to the provided M2 threaded holes (see Figure
above). Pay attention to the specified torque range (p. 23) while doing so.