User Manual
P736T0002, applies to P-736.ZRxS PInano® Piezo Scanner Systems
CBo, 2021-09-20
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Page 13 / 27
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
Mounting the P-736.ZRx onto a Surface
Figure 4: Mounting holes in the base body of the P-736.ZRx
Requirements
You have read and understood the safety precautions (p. 6).
The P-736.ZRx is
not
connected to the controller.
You have provided a suitable surface:
-
The surface has four suitable M3 holes. See “Dimensions” (p. 22) for details.
-
The flatness
of the surface is ≤
10 µm.
-
For applications with large temperature changes: The surface should have the
same thermal expansion properties as the P-736.ZRx (e.g., surface made of
aluminum).
You have accounted for the space required to route cables without bending and
according to regulations.
Tools and accessories
Four M3x10 hex socket head screws (included with P-736.ZRx)
Suitable hex key for M3x10 hex socket head screws
Mounting the P-736.ZRx onto a surface
1.
Align the P-736.ZRx on the surface so that the mounting holes in the P-736.ZRx (see
figure above) are in line with the corresponding holes in the surface.
2.
Insert a mounting screw into each mounting hole in the base body of the P-736.ZRx.
3.
Tighten the mounting screws with a torque between 0.8 Nm and 1.1 Nm.
4.
Check that the P-736.ZRx is affixed firmly to the surface.