5 Installation
26
Version: 1.1.1
PZ245E
P-2x5 Piezo Actuator
Figure 8: P-2x5: Purge air connector with P-177.50 option
1
Inlet for purge air, M3-PK-2 plug nipple
2
Outlet for purge air, M3-PK-2 plug nipple
Requirements
You have read and understood the general notes on installing (p. 19).
The P-2x5 is discharged (p. 34) and short-circuited by the shorting plug (p. 12) provided.
Tools and accessories
Hoses for feeding and discharging the purge air, suitable for M3-PK-2 plug nipple
Suitable purge air:
The requirements on purge air quality can generally be met by running compressed air
conformant to ISO 8573-1:2010.
Particulates: Class 2
Humidity: Class 4 (Temperature at least 3 °C above dew point)
Residual oil content: Class 1 (<0.01 mg/m³, measured at 1 bar and 20 °C)
The pressure in the purge air supply should be between 0.5 bar and 1 bar.
Connecting the purge air
1.
Make sure that the piezo actuator has cooled down to room temperature.
2.
Connect the purge air:
−
Attach the hose for feeding the purge air to the corresponding plug nipple on the
P-2x5 (see above figure).
−
Attach the hose for discharging the purge air to the corresponding plug nipple on
the P-2x5 (see above figure).