10 Technical Data
46
Version: 1.1.0
MP150E
L-406 Compact Linear Stage
10.1.4
Limit Switch Specifications
Type
Optical sensor
Supply voltage
+5 V / ground
Signal output
TTL level
Signal logic
The signal level changes when passing the limit switch. The
signal logic is active high. That means:
Normal motor operation: low (0 V)
Limit switch reached: high (+5 V)
10.1.5
Reference Switch Specifications
Type
Optical sensor
Supply voltage
+5 V / GND, supply via the motor connector
Signal output
TTL level
Signal logic
Direction sensing by means of different signal levels on the
left- and right-hand side of the reference switch: The signal
level changes from 0 to +5 V when the reference switch is
passed.
10.2
Dimensions
10.2.1
L-406.x0DD10 and L-406.x0DG10
Dimensions in mm. Note that a comma is used in the drawings instead of a decimal point.
See "Hole Pattern of the Platform of the L-406" (p. 49) for a detailed view of the platform.