6 Start-Up
20
Version: 1.2.0
MP164E
L-402 Mniature Stage
NOTICE
Damage or considerable wear due to high accelerations!
High accelerations can cause damage to or considerable wear on the mechanical system.
Stop the motion immediately if a controller malfunction occurs.
Ensure that the end of the travel range is approached at low velocity.
Determine the maximum velocity for your application.
NOTICE
Unintentional motions!
Unintentional motion of the L-402 is possible when it is connected to the controller. Defective
software or incorrect operation of the software can also result in unintentional motion.
Do not place any objects in areas where they can be caught by moving parts.
Before connecting the L-402, check whether a macro is defined as the start-up macro in
the controller, and cancel the selection of the start-up macro if necessary.
6.2
Starting up the L-402
Requirements
You have read and understood the general notes on start-up (p. 19).
For starting up with a load or in a multi-axis system: You have properly installed the L-
402 (p. 12).
The L-402 has been properly connected to the controller (p. 18).
You have read and understood the user manual of the controller used.
The required PC software has been installed.
You have read and understood the manual of the PC software used.
Starting up the L-402
1.
Start up the controller (see user manual of the controller).
Configure the controller(s) during start-up using the PC software for the L-402 used (see
user manual of the controller, and the PC software): Select the entry in the stage
database that exactly matches the L-402 model used. (p. 5).
2.
Start a few motion cycles for testing purposes (see user manual of the controller).